![Fluidigm CyTOF XT User Manual Download Page 101](http://html1.mh-extra.com/html/fluidigm/cytof-xt/cytof-xt_user-manual_2308832101.webp)
Chapter 4: Maintenance
Autosampler Module Maintenance
CyTOF XT User Guide
101
Clean the Bottle Tray
Clean the bottle tray monthly or following any spillage.
1
Remove the bottles from the bottle tray.
IMPORTANT
To prevent cross-contamination, do not allow lines from the bottle caps
to touch any fluid in the bottle tray. Let them hang in front of the system.
2
Remove the bottle tray and rinse with tap water. Dry with a Kimwipe.
3
Reinstall the bottle tray.
4
Reattach the caps and reinstall bottles.
NOTE
Wipe the tubing from each cap with a separate Kimwipe moistened with DIW
before putting it back into the bottle.