Chapter 4: Maintenance
Clean the Injector and Spray Chamber
CyTOF XT User Guide
63
2
Gently pull to remove the injector.
Clean the Injector
1
Soak the injector in 10% Contrad/Decon for 1 hr.
IMPORTANT
If injector tip is chipped or broken do not use. Replace the injector.
2
Scrub the injector with the glassware brush provided.
3
Rinse thoroughly with DIW.
4
Spray isopropanol inside the injector to hasten drying. Allow parts to dry thoroughly at
room temperature.
5
Inspect the O-ring on the injector cap for damage or discoloration. Replace the O-ring if
necessary.
Summary of Contents for CyTOF XT
Page 1: ...FLDM 00254 Rev 03...