Chapter 4: Maintenance
Clean the Autosampler Fluidic Lines and Nebulizer with 10% Bleach
86
CyTOF XT User Guide
b
Under
Settings
, set
Flow Rate (µL/min)
to
50
.
c
Under
Sample Settings
, set
Sample Location
to
1
.
d
Click
OK
to close the Flow Rate notification.
e
Click
to save the changes.
8
To load the sample loop with bleach solution, under
Actions > Workflows > Sample
Prepare
, click
Load
.