5.24 Auxiliary Power Output
Voltage:
24VDC regulated and filtered
Isolation:
230VAC max
Current:
0 to 100mA
Protection:
Short Circuit Proof
5.3 RS232 Communications Port
(See Section 8 for more information on RS232 Option.)
Connector:
25 Pin Female (DB25S)
Input Impedance:
3000
Ω
to 7000
Ω
Compliance Voltage:
Output: -25 to -5 (Mark); 5 to 25 (Space); Volts
Input: -25 to -3 (Mark); 3 to 25 (Space); Volts
Protection:
Short circuit proof.
Protocol:
8 bits, 1 Stop bit
Parity:
None (Not monitored)
Available Baud Rates: 300, 1200, 9600
RS232 Connector Pin Out
5.4 Data Display and Keypad
Internal 2 line by 20 character dot matrix LCD dis-
play. Sealed, 16 key panel featuring numeric keys
0- 9, plus the following keys:
A ...... Advance through menus
B ...... Back up through menus
C ...... Cancel current menu selection
D ...... Decimal point key
ENT . General purpose enter or recall data key
CLR . Data clear key
5-6
Ground (7)
Transmit (3)
Printer Busy (11)
Receive (2)
13 12 11 10 9 8 7 6 5 4 3 2 1
DB25S
25 24 23 22 21 20 19 18 17 16 15 14
Pins 6, 8 and 20 are internally jumpered together
Pins 4 and 5 are internally jumpered together