3.12 Setting the Real Time Clock
The instrument has a real time clock/calendar that can be
set from the front keypad.
The time and date are saved upon power down. However,
there is no battery back up in the Flow Computer to up-
date the time and date during loss of power. For this
reason, if you are using this feature, the time and date
must be reset upon each power up.
To do this, follow the menu listed below:
DISPLAY
PRESS
Run?
A
Set Computations?
A
Continue Pressing A until
Check Alarm?
A
Real time clock?
ENT
Select Item:
Set Time?
ENT
Enter time:
hr.min ?
i.e. 14.32
Enter time:
hr.min ? 14.32
ENT
Remember that when entering the time for the real
time clock, it must be in 24 hour format.
Set Time:
Hours (00 to 23) and minutes (00 to 59) based on 24 hour
clock.
3-26