4-4
Freq03? 20.0000
K Factor03?
1.11111
Enter Point
Freq04?
30
Freq04? 30.0000
K Factor04?
1.017
Enter Point
Freq05?
100
Freq05? 100.0000
K Factor05?
1.000
Enter Point
Freq06?
1000
Freq06? 1000.0000
K Factor06?
1.000
Note: Since the K factors for the last two points are one,
any frequency higher than 1000 Hz will be modified by a
K factor of 1.
4.3 16 Point Analog Setup
Analog programming is same as the Digital setup. The
Flow Computer asks for 16 different flow rates and 16
corresponding K factors (correction factors). It then uses
this data to determine what the true flow is for any given
actual (apparent) input signal. Again, a minimum of
three points must be set up.
K factor =
Actual Input Flow
True Flow
The same rules that applied for the Digital setup apply for
the Analog setup as well.