Step Ten
The
Solid Medium
function has the following settings (
and should not be performed without first consulting
with the factory
):
Fast Level Change
- Used when the media’s level
rise or fall within the tank is greater than a rate of 1”
per second (25.4mm/sec). Selections are YES or
NO.
Default is NO
.
First Echo
- Used when the sensor has difficulty
seeing the first echo return. This setting adjusts the
peak strength (dB) of the first echo.
Default is
Normal
. Below are the settings:
o
Normal
- No adjustment
o
Small
-Decreases by 10dB
o
Big
- Increases by 10dB
o
Bigger
- Increases by 20dB
o
Biggest
- Increases by 40dB
Large Repose Angle -
Used when the repose angle
of the material is steep. Repose angle is the
steepest angle of assent or descent relative to the
horizontal plane that a material can be piled without
sliding. The repose angle varies between different
materials. Selections are Yes or No.
Default is No
.
Powder / Dust -
Used when powder or dust are
present in the atmosphere above the material.
Selections are Yes or No.
Default is No
.
Low Dielectric
- Used when the material has a low
dielectric constant (
and should not be performed
without first consulting with the factory
). Selections
are YES or NO.
Default is NO
.