40
MN301905 Rev A
Configuration
(with WebCal®)
Step Six
BASIC CONFIGURATION OVERVIEW
Below are the 7 basic steps to configure the sensor for operation. Each step is described in detail on the
following pages
1) Install WebCal
®
software
a) Go to
www.flowline.com/webCal-software/
and select language version.
b) Review how USB
®
Radar Fob interfaces with EchoPro
®
and your computer.
2) Measure the Tank
a) Begin by measuring the key tank and fitting dimensions. Correct tank dimensions will result in accurate
sensor measurement.
3) Sensor Configuration
a) Configures Loop Fail-Safe, Output at Empty, Minimum Current & Display for the sensor.
4) Dimensional Entry
a) Distance Mode (default)
i) Basic information for 4-20 mA operation (Sensor Height & Fill-Height).
ii) Basic information for operational range operation (Maximum Range & Dead Band).
5) Tank Level Confirmation
a) Confirm the values are accurate for the application.
6) Write to Unit
a) Uploads configuration into the sensor.
b) Access to a customer wiring diagram specific to the relay configuration.
7) Check EchoCurve