MN301905 Rev A
Troubleshooting (continued)
Step Nine
RESET
This function resets the sensor’s configuration (basic or factory) settings and memory (peak values measured).
1) From the Main Screen, press
Enter
to advance into the
Main Menu.
2)
Press
Right Arrow
repeatedly until the arrow is next to
Service.
3)
Press
Enter
to advance into Service.
4)
Press
Right Arrow
repeatedly until Reset appears.
5)
Press
Enter
to advance into Reset.
6)
Use the
Right Arrow
to scroll among the three reset types
(Basic Reset, Factory Settings, Peak Value Meas).
a)
Basic Reset
i)
Resets basic configuration settings, process adjustment settings, and peak level values
in memory (retaining any advanced adjustment settings).
b)
Factory Settings
i)
Resets ALL settings and memory to factory default.
Note: Only perform the Factory Setting after consulting with a Flowline Representative. Make sure to record
all of your settings including the Distance Adjustment (4.9) setting before performing the Factory Reset.
c)
Peak Values Measured
i)
Resets the minimum and maximum peak level values in memory (retaining any basic
configuration, process adjustment and advanced adjustment settings).
7) Use
Enter
to select the desired type of reset.
8) When done, press
ESC
to return to Service, and press
ESC
a second time to the Main Screen.