37
7.3.3.4
Fixing the rail to the pipe
• Place one tension strap clamp in the rail (see tension strap clamp 1 in Fig. 7.20). Observe the orientation of the tension
strap clamp.
• Tighten the nut of tension strap clamp 1 slightly.
• Screw the rail to tension strap clamp 2 (see Fig. 7.21).
• Tighten the nut of tension strap clamp 2, but not too firmly in order not to damage the tension strap.
• Select the installation instructions that correspond to the supplied clasp:
Band clamp clasp
• Insert the tension strap into tension strap clamp 1 (see Fig. 7.22).
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 7.23).
• Tighten the tension strap.
• Tighten the screw of the clasp.
• Tighten the nut of tension strap clamp 1, but not too firmly in order not to damage the tension strap (see Fig. 7.22).
Fig. 7.20: Rail with tension strap clamp
Fig. 7.21: Rail, mounted on one side of the pipe
Fig. 7.22: Rail on the pipe
Observe the
orientation!
nut
tension strap 1
nut
tension strap 2
tension strap 2
tension strap 1
nut
clasp
metal spring
tension strap 1