6 Measuring point selection
24
Observe the notes in the following table:
Tab. 6.1: Recommended transducer mounting position
horizontal pipe
Select a measuring point where the transducers can be mounted on the side of the pipe, allowing the sound waves to
propagate horizontally in the pipe. Thus, solids on the bottom of the pipe or gas bubbles in the pipe's upper part are pre-
vented from influencing the propagation of the signal.
correct:
disadvantageous:
vertical pipe
Select the measuring point in such way that the fluid flows upward. The pipe has to be completely filled.
correct:
disadvantageous:
free inlet or outlet pipe section
Select the measuring point at a pipe section where the pipe cannot run empty.
correct:
disadvantageous:
correct:
disadvantageous: