11 Displaying measured values
81
11.5
Transducer distance
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses in the upper line, followed by the en-
tered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during the measurement (e.g., due to temperature fluctuations).
A deviation from the optimum transducer distance (here: -0.4 mm) will be compensated internally.
By pressing key
during the measurement, it is possible to scroll to the display of the
transducer distance.
Note!
Never change the transducer distance during the measurement!
L=(51.2) 50.8 mm
54.5
m3/h