15 SuperUser mode
105
15
SuperUser mode
The SuperUser mode offers the possibility of an advanced diagnostic of signal and the measured values as well as the
definition of additional parameters adapted to the measuring point, in order to achieve better measuring values or for ex-
perimental work. Special features of the SuperUser mode are:
• Default settings will not be used.
• There are no plausibility tests during the parameter entry.
• It is not checked whether the entered parameters are within the limits which are determined by physical laws and techni-
cal data.
• The cut-off flow is not active.
• A value for the number of sound paths has to be entered.
• Some menu items that are not visible in the normal the normal mode are displayed.
15.1
Activation/deactivation
Press key C. Enter the HotCode
071049
.
Press key C. Enter the HotCode
071049
again to deactivate the SuperUser mode.
15.2
Transducer parameters
In the SuperUser mode, the menu item
Transducer Type
will be displayed in the program branch
Parameter
at the
end of the input even if the transducers are detected by the transmitter.
15.3
Defining flow parameters
In the SuperUser mode, it is possible to define some flow parameters (profile bounds, correction of the flow velocity) for
the specific application or measuring point.
Attention!
The SuperUser mode is intended for experienced users with advanced application knowledge. The
parameters can affect the normal measuring mode and lead to wrong measuring values or to a mea-
surement failure when a new measuring point is set up.
It is displayed that the SuperUser mode is activated. Press ENTER.
The main menu will be displayed.
It is displayed that the SuperUser mode is deactivated. Press ENTER.
The main menu will be displayed.
Attention!
Some of the defined parameters are still active after the deactivation of the SuperUser mode.
Press ENTER.
or:
Select
Special Version
to enter the transducer parameters.
Press ENTER.
If
Special Version
is selected, the transducer parameters have to be entered.
The transducer parameters have to be provided by the manufacturer. Press ENTER after
each input.
Select
Special Funct.
\
SYSTEM settings\Measuring\Calibration
.
Press ENTER.
Select the measuring channel for which the flow parameters are to be defined.
Press ENTER.
SUPERUSER MODE
*IS ACTIVE NOW*
SUPERUSER MODE
IS PASSIVE NOW
Transducer Type
↕
Q2E-314
Transducer Type
↕
Special Version
Transd. Data
1
35.99
Measuring
↕
Calibration
Calibrat. data
↕
for Channel
A:
Summary of Contents for FLUXUS WW
Page 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Page 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Page 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Page 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Page 145: ......
Page 146: ......