A Menu structure
132
Special Funct.\SYSTEM settings\Measuring\Calibrat. data
selection of the measuring channel whose flow parameters are to be de-
fined
This display will only be indicated if the SuperUser mode is activated.
determination of the profile bounds
factory
: the default profile bounds will be used
user
: the profile bounds can be defined
This display will only be indicated if the SuperUser mode is activated.
input of the max Reynolds number at which the flow is laminar
range: 0...25 500 (rounded to hundreds)
0: the default limit of 1 000 will be used
This display will only be indicated if the SuperUser mode is activated and
Profile bounds
=
user
is selected.
input of the min. Reynolds number at which the flow is turbulent
range: 0...25 500 (rounded to hundreds)
0: the default limit of 3 000 will be used
This display will only be indicated if the SuperUser mode is activated and
Profile bounds
=
user
is selected.
request if an additional correction of the flow velocity is to be defined
on
: the correction data can be defined
off
: without correction of the flow velocity
This display will only be indicated if the SuperUser mode is activated.
input of the slope for the correction formula
range: -2.000...+2.000
0: without correction
This display will only be indicated if the SuperUser mode is activated and
Calibration
=
on
is selected.
input of the offset
range: -12.7…+12.7 cm/s
0: without offset
This display will only be indicated if the SuperUser mode is activated and
Calibration
=
on
is selected.
SYSTEM settings\Proc. outputs
selection of the displays for the transmitter outputs setting
selection of the output to be installed
SYSTEM settings\Storing
selection of the displays for the storing of measured values in the data
logger
setting of the data logger overflow behavior
x
INIT-resis-
tant
Calibrat. data
↕
for Channel
A:
A:Profile bounds
factory
>USER<
Laminar flow
ifR*<
0
Turbulent flow
if R*>
0
A:Calibration ?
>OFF<
on
A:Slope=
1.000
A:Offset=
0.0
cm/s
SYSTEM settings
↕
Proc. outputs
Install Output
↕
Current I1
SYSTEM settings
↕
Storing
Ringbuffer
off
>ON<
Summary of Contents for FLUXUS WW
Page 1: ...User manual UMFLUXUS_WWV1 1EN Ultrasonic flowmeter for water and wastewater FLUXUS WW...
Page 4: ...FLUXUS WW UMFLUXUS_WWV1 1EN 2018 02 28...
Page 8: ...FLUXUS WW Table of contents 8 UMFLUXUS_WWV1 1EN 2018 02 28...
Page 144: ...FLUXUS WW D Conformity declarations 144 UMFLUXUS_WWV1 1EN 2018 02 28 D Conformity declarations...
Page 145: ......
Page 146: ......