9 Measurement
9.5 Execution of special functions
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
108
9.5
Execution of special functions
Commands that can be executed during a measurement are displayed in the upper line.
A command begins with →. If programmed, a program code has to be entered first.
• Press key
until the command is displayed.
• Press ENTER.
The commands available are the following:
9.6
Determination of the flow direction
The flow direction in the pipe can be detected with the help of the displayed volumetric flow rate sign in conjunction with
the arrow on the transducers:
• The fluid flows in the direction of the arrow if the displayed volumetric flow rate is positive (e.g., 54.5 m³/h).
• The fluid flows against the arrow direction if the displayed volumetric flow rate is negative (e.g., -54.5 m³/h).
9.7
Stop of the measurement
The measurement is interrupted by pressing key BRK if it is not protected by a program code, see section 18.6.
Tab. 9.6:
Commands that can be executed during the measurement
command
explanation
→Adjust transd.
toggle to transducer positioning
If a program code is activated, the measurement will be continued 8 s after the last keyboard entry.
→Clear totalizer
reset of flow totalizers to zero
→Mux:Auto/Human
toggle between the AutoMux and the HumanMux mode of the display
This display will not be indicated if the transmitter has only one measuring channel or only one mea
-
suring channel is activated.
→Mux:Nextchan.
display the next channel
This display will not be indicated if the transmitter has only one measuring channel or only one mea
-
suring channel is activated.
→Break measure
stop the measurement and return to the main menu
→Toggle FastFood
Notice!
Be careful not to stop a current measurement by inadvertently pressing key BRK.
A:Mode=FastFood
54.5
m3/h
A:Mode=TransTime
54.5
m3/h