13 Outputs
13.2 Configuration of a frequency output as pulse output
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
120
Function test of the binary output
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test the de-energized state of the
output.
• Press ENTER. Measure the resistance at the output. The value has to be high ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the energized state of the output.
• Press ENTER. Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
13.2
Configuration of a frequency output as pulse output
A frequency output sends a signal with a frequency that depends on the volumetric flow rate. The frequency output can be
configured in such way that the source item can be totalized by using each period of the output signal as the increment.
Installation of a frequency output (optional)
• Select the frequency output to be installed.
• Press ENTER.
• Select
yes
to install or reconfigure the output.
• Press ENTER.
• Select
no
to uninstall the output and to return to the previous menu item in order to select another output.
• Press ENTER.
Tab. 13.1: Configuration of the binary outputs
source item
list item
output
Limit
R1
limit message (
Alarm Output
R1
)
R2
limit message (
Alarm Output
R2
)
R3
limit message (
Alarm Output
R3
)
Impuls
from abs(x)
pulse without sign consideration
from x > 0
pulse for positive measured values of the volumetric flow rate
from x < 0
pulse for negative measured values for the volumetric flow rate
Special Funct.\SYSTEM settings\Proc. outputs\...\Output Test B1\Reed-Relay OFF
Special Funct.\...\Output Test B1\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\...\Output Test B1\Reed-Relay ON
Special Funct.\...\Output Test B1\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\Install Output
Special Funct.\SYSTEM settings\Proc. outputs\Enable F1