14 Data logger
14.3 Configuration of the data logger
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
124
14.3
Configuration of the data logger
• Select
Special Funct.\SYSTEM settings\Storing
.
• Press ENTER.
Starting time
In order to start the storing of measured values with several measuring instruments at the same time, it is possible to set a
starting time.
• Select the moment at which the storing has to start.
Ringbuffer
The setting of the ringbuffer influences the storing of measured values as soon as the data logger is full:
• If the ringbuffer is activated, the available data logger will be halved. The oldest measured values will be overwritten.
Only the data logger memory that was free during the activation will be used by the ringbuffer. If more data logger mem
-
ory is necessary, measured values in the data logger should be previously deleted.
• If the ringbuffer is deactivated, the storing of measured values will be stopped.
• Select
ON
to activate the ringbuffer.
• Press ENTER.
Storage mode
• Select the storage mode.
• Press ENTER.
If
sample
is selected, the current measured value will be used for the storing and online transmission of data.
If
average
is selected, the average of all values measured during a storage interval will be used for the storing and online
transmission of data.
Special Funct.\SYSTEM settings\Storing
Special Funct.\SYSTEM settings\Storing\Start logger
display
description
Promptly
The storing starts immediately.
On full 5 min.
The storing starts in the next full 5 minutes.
On full 10 min.
The storing starts in the next full 10 minutes.
On quarter hour
The storing starts in the next full 15 minutes.
On half hour
The storing starts in the next full 30 minutes.
On full hour
The storing starts in the next full 60 minutes.
Example
actual time: 09:06 am
setting:
On full 10 min.
The storing starts at 09:10 am.
Special Funct.\SYSTEM settings\Storing\Ringbuffer
Special Funct.\SYSTEM settings\Storing\Storage mode