7 Connection
7.1 Transducers
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
66
7.1.3
Connection of the transducer cable to the junction box
7.1.3.1
Transducer cable with plastic cable jacket and stripped cable ends
• Remove the blind plug for the connection of the transducer cable, see Fig. 7.6.
• Open the cable gland of the transducer cable. The compression part remains in the cap nut.
• Push the transducer cable through the cap nut and the compression part.
• Prepare the transducer cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the gasket ring side of the basic part into the junction box.
• Insert the transducer cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the transducer cable to the terminals of the junction box, see Fig. 7.6 and Tab. 7.4.
Fig. 7.5:
Connection of the extension cable to the transmitter
1 – external shield brushed back
Notice!
For good electromagnetic compatibility (EMC), it is important to ensure good electrical contact between the external
shield and the cap nut (and thus the housing).
Tab. 7.4:
Terminal assignment
terminal
connection
V
transducer (core)
VS
transducer (internal shield)
RS
transducer (inner shield)
R
transducer (core)
AR ARS
AV
AVS
BR BRS
BV
BVS
70 mm
20 mm
10 mm