9 Measurement
FLUXUS G809
9.2 Measurement settings
97
UMFLUXUS_G809V2-2EN, 2020-06-25
9.2.3.1
Output of the measured value
Output range
When configuring an analog output, the output range has
to be defined.
• Select the menu item
Special Funct.\
SYSTEM settings\Proc. outputs\...\Output
range I1
.
• Press ENTER.
• Select a list item.
– 4/20 mA
– other range...
• Press ENTER.
• If
other range
is selected, enter the values
Output MIN
and
Output MAX
.
• Press ENTER after each input.
Error output
In the following dialog, an error value can be defined which is to be transmitted if the source item cannot be measured,
e.g., if there are solids in the fluid.
• Select a list item for the error output, see Tab. 9.2.
• Press ENTER.
• If
Other value...
is selected, enter an error value. The value has to be within the output range.
• Press ENTER.
Source item
Output range
Error-value
Active loop
Output Test
Special Funct.\SYSTEM settings\
Proc. outputs\...\Output range I1
Special Funct.\SYSTEM settings\Proc. outputs\...\Error-value I1
Notice!
The settings will be stored at the end of the dialog.
Tab. 9.2:
Error output
error value
result
Minimum
the lower limit of the output range is output
Hold last value
the last measured value is output
Maximum
the upper limit of the output range is output
Other value...
The value has to be entered manually. It has to be within the limits of the output.