18 Settings
18.6 Program code
FLUXUS G809
2020-06-25, UMFLUXUS_G809V2-2EN
164
18.6
Program code
A running measurement can be protected against inadvertent intervention by a program code.
If a program code has been defined, it will be requested in case of any intervention in the measurement (a command or
key BRK).
18.6.1
Defining a program code
• Select
Special Funct.\Program code
.
• Press ENTER.
• Enter a program code with max. 6 digits.
• Press ENTER.
An error message will be displayed if a reserved number has been entered (e.g., a HotCode for language selection).
A program code will remain valid as long as:
• no other valid program code is entered or
• the program code is not deactivated
18.6.2
Intervention in the measurement
If a program code is activated, the message
PROGRAM CODE ACTIVE
will be displayed for a few seconds when a key is
pressed.
The input of a program code is interrupted by pressing key CLR.
If key BRK is pressed:
In order to stop a running measurement, the complete program code has to be entered (= Break Code).
• Enter the program code with the keys
and
.
• Press ENTER.
If the entered program code is invalid, an error message will be displayed for a few seconds.
If the entered program code is valid, the measurement will be stopped.
Special Funct.\Program code
Notice!
Do not forget the program code!
INVALID CODE !
909049
INPUT BREAK_CODE
CODE:
000000
INPUT BREAK_CODE
INVALID CODE !