13 Advanced Measuring Functions
UMFLUXUS_G8V4-6-1EN, 2018-10-10
95
13.10
Diagnosis with the Snap Function
With the aid of the snap function it is possible to store measuring parameters which are useful for the evaluation of mea-
suring results or for diagnostic purposes.
The snap function is activated in
Special Funct.
\
SYSTEM settings
\
Signal snap
\
DSP-SignalSnap
.
Settings of the snap memory
Delete Snaps
Read Snaps
Activation of the Snap Function
In order to activate the snap function, press key
during the measurement until
DSP-SignalSnap/Voltage
is dis-
played in the upper line. Press ENTER.
13.11
Program Code
An ongoing measurement can be protected from an inadvertent intervention by means of a program code.
If a program code has been defined, it will be requested when there is an intervention in the measurement (a command or
key BRK).
13.11.1 Defining a Program Code
A program code will remain valid as long as:
• no other valid program code is entered or
• the program code is not deactivated.
Select
on
to activate the snap function. Select
off
to deactivate the snap function.
Select
Install Snap
. Press ENTER.
Enter the number of the snap memory capacity. Press ENTER.
Activation and deactivation of the auto-snap function. Press ENTER.
Activation and deactivation of the snap ringbuffer. Press ENTER.
Select
Clear Snaps
. Press ENTER.
Select
Snaps ->Rs232
. Press ENTER.
Select
Special Funct.\Program code
.
Enter a program code with max. 6 digits. Press ENTER.
An error message will be displayed if a reserved number has been entered (e.g., a Hot-
Code for language selection).
DSP-SignalSnap
off
>ON<
DSP-SignalSnap
Install Snap
Snap-Memory
5
NUM
AutoSnap
>NO<
yes
Snap ringbuffer
>NO<
yes
DSP-SignalSnap
Clear Snaps
DSP-SignalSnap
Snaps ->Rs232
Special Funct.
↕
Program code
Program code
------
INVALID CODE !
909049