A Menu Structure
138
UMFLUXUS_G8V4-6-1EN, 2018-10-10
selection of the sound speed range
auto
: The area around the average sound speed is defined by the transmit-
ter.
user
: The area around the average sound speed has to be entered.
input of the area around the average sound speed of the fluid
This display will only be indicated if
user
is selected.
input of the kinematic viscosity of the fluid
range: 0.01...30 000 mm²/s
This display will only be indicated if
Other Medium
is selected or no data set
for the selected fluid is stored in the transmitter.
input of the operation density of the fluid
range: 0.1...20 000 kg/m³, if
Special Funct.
\
SYSTEM settings
\
Mea-
suring
\
Gas-Measuring
is activated or 0.01...20 g/cm³ if
Special
Funct.
\
SYSTEM settings\Measuring
\
Gas-Measuring
is deactivated
This display will only be indicated if
Other Medium
is selected or no data set
for the selected fluid is stored in the transmitter.
input of the compressibility coefficient of the gas
range: 0.001...2
This display will only be indicated if
Other Medium
is selected or no data set
for the selected fluid is stored in the transmitter.
input of the fluid temperature
default: 20 °C
input of the fluid pressure
range: 1...600 bar
This display will only be indicated if
Special Funct.\SYSTEM set-
tings\Measuring\Gas-Measuring
is activated or if
Gas-Measuring
is
deactivated and \
Special Funct.
\
SYSTEM settings
\
Dialogs/Menus
is activated.
selection of the transducer type
This display will only be indicated if no or special transducers are connected.
input of the length of an extension cable
When a calculation channel is selected (Y, Z)
Calculation channels are only available if the transmitter has more than one measuring channel.
display of the current calculation function
selection of the calculation function
Program branch
Measuring
main menu: selection of the program branch
Measuring
INIT-resis-
tant
c-Medium range
auto
>USER<
c-Medium=400m/s
range +-150m/s
Kinem.Viscosity
1.00
mm2/s
Density
60.00 kg/m3
Gas compr.factor
1.000 factor
Medium Temperat.
20.0
C
Fluid pressure
60.00
bar
Transducer Type
↕
Standard
Additional cable
65.0
m
Calculation:
Y= A - B
>CH1< funct ch2
↕
A - B
par >MEA< opt sf
Measuring