11 Basic Measurement
80
UMFLUXUS_G8V4-6-1EN, 2018-10-10
11.2.6
Fluid Pressure
The fluid pressure is used for the interpolation of the sound speed.and the compressibility coefficient of the gas
11.3
Other Parameters
11.3.1
Transducer Parameters
If transducers are detected at a measuring channel, the input of parameters will be finished. Press ENTER. The main
menu will be displayed.
If no or special transducers are connected, the transducer parameters have to be entered.
11.3.2
Extension Cable
11.4
Selection of the Channels
The channels on which the measurement is conducted can be activated individually.
• Select a channel with the key
.
• Press key
to activate or deactivate the selected channel. Press ENTER.
A deactivated channel will be ignored during the measurement. The parameters entered for this channel will remain un-
changed.
Enter the fluid pressure. Press ENTER.
This display will only be indicated if
Special Funct.
\
SYSTEM settings
\
Measur-
ing
\
Gas-Measuring
is activated OR if
Gas-Measuring
is deactivated and
Special
Funct.
\
SYSTEM settings
\
Dialogs/Menus
\
Fluid pressure
is activated.
Select
Standard
to use the standard transducer parameters stored in the transmitter.
Select
Special Version
to enter the transducer parameters. The transducer parame-
ters have to be provided by the manufacturer.
Press ENTER.
Note!
If standard transducer parameters are used, FLEXIM cannot guarantee for the precision of the mea-
sured values. A measurement might even be impossible.
If
Special Version
is selected, enter the 6 transducer parameters specified by the
manufacturer. Press ENTER after each input.
If the transducer cable is extended, enter the additional cable length (e.g., between the
junction box and the transmitter). Press ENTER.
Select the program branch
Measuring
. Press ENTER.
If this error message is displayed, the parameters are not complete. Enter the missing pa-
rameters in the program branch
Parameter
.
The channels for the measurement can be activated and deactivated.
: the channel is active
–: the channel is not active
•: the channel cannot be activated
This display will not be indicated if the transmitter has only one measuring channel.
Note!
A channel cannot be activated if the parameters are not valid, e.g., if the parameters in the program
branch
Parameter
of the channel are not complete.
Fluid pressure
60.00
bar
Transducer Type
↕
Standard
Transd. Data
1
35.99
Additional cable
65.0
m
par>MEA<opt sf
Measuring
par>MEA<opt sf
NO DATA!
CHANN:
MEASUR
< >A< B Y Z
- .