A Menu Structure
UMFLUXUS_G8V4-6-1EN, 2018-10-10
143
selection if lb/ft³ is to be used as unit of measurement for the density
x
selection of the unit of measurement for the density
This display will only be indicated if lb/ft³ has not been selected as the unit of
measurement for the density.
x
selection of the unit of measurement for the kinematic viscosity
x
selection of the unit of measurement for the sound speed
x
SYSTEM settings\Measuring
selection of the displays for the measurement settings
activation of the gas measurement
default:
on
x
input of an upper limit of the flow velocity
range: 0.1...25.5 m/s
0 m/s: no detection for outliers
All measured values that are greater than the limit will be marked as outliers.
x
selection of the input of a lower limit for the flow velocity:
•
absolut
: independent of the flow direction
•
sign
: dependent on the flow direction
x
activation of the input of a lower limit of the flow velocity:
•
factory
: the default limit of 2.5 cm/s will be used
•
user
: input of a limit
x
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s (0.127 m/s)
default: 2.5 cm/s (0.025 m/s)
This display will only be indicated if
Cut-off Flow
=
sign
and
Cut-off
Flow
=
user
is selected.
x
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if
Cut-off Flow
=
sign
and
Cut-off
Flow
=
user
is selected.
x
input of the cut-off flow for the absolute value of the measured values
range: 0…12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if
Cut-off Flow
=
absolut
and
Cut-
off Flow
=
user
is selected.
x
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
INIT-resis-
tant
Density [lb/ft3]
no
>YES<
Density unit
g/cm3
>kg/m3<
Viscosity unit
mm2/s
>cSt<
Soundspeed unit
>[m/s]<
[fps]
SYSTEM settings
↕
Measuring
Gas-Measuring
off
>ON<
Velocity limit
0.0
m/s
Cut-off Flow
absolut
>SIGN<
Cut-off Flow
factory
>USER<
+Cut-off Flow
2.5
cm/s
-Cut-off Flow
-2.5
cm/s
Cut-off Flow
2.5
cm/s
A: Gain threshold
Fail if>
90