14 Data Logger and Transmission of Data
UMFLUXUS_G8V4-6-1EN, 2018-10-10
101
14.2.4
Transmission Parameters
• the transmitter sends CRLF-terminated ASCII
• max. line length: 255 digits
RS-232
• default: 9600 bits/s, 8 data bits, even parity, 2 stop bits, protocol RTS/CTS (hardware, handshake)
The transmission parameters of the RS232 interface can be changed:
RS485
• default: 9600 bits/s, 8 data bits, even parity, 1 stop bit
The transmission parameters of the RS485 interface can be changed in the program branch
Special Funct.
\
SYSTEM
settings
\
Network
. This display will only be indicated if the transmitter has an RS485 interface.
Select the decimal marker to be used for floating-point numbers (point or comma). Press
ENTER.
This setting depends on the setting of the operating system of the PC.
Select the character to be used to separate columns (semicolon or tabulator). Press EN-
TER.
Enter HotCode
232-0
Set the transmission parameters in the 4 scroll lists. Press ENTER.
•
baud
: baud rate
•
data
: number of data bits
•
par
: parity
•
st
: number of stop bits
Select
Special Funct.\SYSTEM settings\Network
to change the settings of the
transmission parameters.
Press ENTER to confirm the address of the measuring instrument in the network.
Select
default
to display the default transmission parameters.
Select
setup
to change the transmission parameters. Press ENTER.
Set the transmission parameters in the 3 scroll lists. Press ENTER.
•
baud
: baud rate
•
parity
: parity
•
st
: number of stop bits
The default transmission parameters will be set if
default
is selected and the transmis-
sion parameters have not been changed.
SER:decimalpoint
’.’
>’,’<
SER:col-separat.
’;’
>’TAB’<
BAUD<data par st
9600 8bit EVEN 2
SYSTEM settings
↕
Network
Device address:
0 ADR
RS485 protocol
default
>SETUP<
>BAUD< parity st
9600 EVEN 1