User Manual

UMFLUXUS_G8V4-6-1EN

Ultrasonic Flowmeter 

for in Explosive Atmospheres

FLUXUS G800

FLUXUS G801

Summary of Contents for FLUXUS G800

Page 1: ...User Manual UMFLUXUS_G8V4 6 1EN Ultrasonic Flowmeter for in Explosive Atmospheres FLUXUS G800 FLUXUS G801...

Page 2: ...M GmbH Boxberger Stra e 4 12681 Berlin Germany Tel 49 30 936 67 660 Fax 49 30 936 67 680 E mail flexim flexim de www flexim com User manual for FLUXUS G80x UMFLUXUS_G8V4 6 1EN 2018 10 10 Copyright FLE...

Page 3: ...t 10 5 The transmitter can be operated in the language of your choice see section 10 5 Il est possible de s lectionner la langue utilis e par le transmetteur l cran voir section 10 5 El caudal metro p...

Page 4: ...FLUXUS G80x UMFLUXUS_G8V4 6 1EN 2018 10 10...

Page 5: ...easuring Conditions 25 5 6 Selection of the Sound Beam Plane near an Elbow 26 6 Installation of FLUXUS G800 27 6 1 Location 27 6 2 Opening and Closing the Housing 27 6 3 Installation 27 6 4 Connection...

Page 6: ...lection of the Physical Quantity and of the Unit of Measurement 84 12 2 Toggling between the Channels 85 12 3 Adjustment of the Display 86 12 4 Status Line 86 12 5 Transducer Distance 87 13 Advanced M...

Page 7: ...Sum 122 17 10 Display of the Last Valid Measured Value 122 17 11 Display During the Measurement 122 18 Outputs 123 18 1 Installation of an Output 123 18 2 Error Value Delay 126 18 3 Activation of an A...

Page 8: ...FLUXUS G80x Content 8 UMFLUXUS_G8V4 6 1EN 2018 10 10...

Page 9: ...cified in the sales contract provid ed the measuring instrument is used for the purpose for which it was designed and operated according to the instructions given in this user manual Improper use of t...

Page 10: ...strument it is important to pay special attention to the instructions given in this user manual particularly to the following points Protect the transmitter from shocks The housing may only be opened...

Page 11: ...be fully developed and axisymmetrical The shape of the flow profile depends on whether the flow is laminar or turbulent and is influenced by the conditions in the supply line of the measuring point se...

Page 12: ...ng volumetric flow rate v A The volume of the fluid that passes through the pipe per unit time The operating volumetric flow rate is calculated from the product of the flow velocity v and the cross se...

Page 13: ...than the transit time against the flow direction This transit time difference is proportional to the av erage flow velocity The average flow velocity of the fluid is calculated as follows V kRe ka wi...

Page 14: ...urement is conducted in the diagonal arrangement see Fig 3 5 even if the measurement is conducted in the reflection arrangement see Fig 3 6 Beam The path covered by the ultrasonic signal between the t...

Page 15: ...ransducers Sound beam plane The plane containing one two or more sound paths or beams see Fig 3 9 reflection arrangement diagonal arrangement positive transducer distance diagonal arrangement negative...

Page 16: ...ngement with 1 beam 1 transducer pair 1 sound path 1 beam 1 plane 1 transducer pair 2 sound paths 1 beam 1 plane diagonal arrangement with 2 beam reflection arrangement with 2 beams and 2 planes 2 tra...

Page 17: ...he connection of the transducers are in the lower housing the terminals for the connection of the out puts and of the power supply on the back side of the upper housing see Fig 4 1 4 2 FLUXUS G801 The...

Page 18: ...by inadvertently pressing key BRK Tab 4 2 Navigation scroll to the right or up through a scroll list scroll to the left or down through a scroll list Tab 4 3 Input of digits move the cursor to the ri...

Page 19: ...e ambient temperature at the measuring point has to be within the operating temperature range of the transmitter and the sensor see Technical Specification If the measuring point is within an explosiv...

Page 20: ...ly then it can be assumed that the flow profile in the pipe is fully developed However measuring results can be obtained even if the recommended distance to disturbance sources cannot be observed for...

Page 21: ...diffuser supply line l 60 D return line l 10 D disturbance source valve supply line l 80 D return line l 20 D disturbance source reducer supply line l 20 D return line l 10 D disturbance source compre...

Page 22: ...d by the transducer at the same time Fig 5 1 Propagation of ultrasonic waves Tab 5 3 Measuring points to be avoided Measuring point directly at the reflection point lS 3 D Measuring point at a distanc...

Page 23: ...ainless steel length of pipe segment 1 20 D length of pipe segment 2 20 D number of sound paths 2 area with developed flow profile disturbance source 90 elbow recommended area for the measuring point...

Page 24: ...ance source 90 elbow recommended area for the measuring point l 20 D complete pipe segment 2 see Tab 5 2 area with low influence of noise reflection point flange recommended area for the measuring poi...

Page 25: ...eed range compared to the diagonal arrangement transverse flow effects are compensated for because the beam crosses the pipe in 2 directions higher accuracy of measurement because the accuracy increas...

Page 26: ...o the elbow plane With measurements in both directions With measurements in the reflection arrangement with 2 beams and 2 planes The sound beam plane see section 3 3 1 is selected according to the nea...

Page 27: ...ng to the instrument mounting plate 3 see Fig 6 1 Fix the transmitter to the wall 6 3 2 Pipe Installation Installation on a 2 pipe Fix the pipe mounting plate 2 to the pipe see Fig 6 1 Fix the instrum...

Page 28: ...annel B transducers measuring channel A equipotential bonding terminal lower housing equipotential bonding terminal upper housing power supply 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 PE N L L1 L 1 2 5 6 L...

Page 29: ...ough the cap nut the compression part and the basic part of the cable gland Prepare the extension cable Press the cap nut with the compression part on the cable until the thin rim of the compression p...

Page 30: ...e extension cable If earthing on the same potential cannot be ensured see section 6 5 2 2 Remove the left blind plug for the connection of extension cable see Fig 6 7 Open the cable gland of the exten...

Page 31: ...2 Fig 6 7 Connection of the extension and transducer cable to the junction box JB01 Attention For good high frequency shielding it is important to ensure good electrical contact between the exter nal...

Page 32: ...nsion cable The compression part remains in the cap nut Push the extension cable through the cap nut the compression part and the basic part Tab 6 3 Terminal assignment transducer cable terminal conne...

Page 33: ...the gasket ring side of the basic part into the upper housing of the transmitter Fix the cable gland by screwing the cap nut onto the basic part see Fig 6 9 Connect the leads to the terminals of the t...

Page 34: ...ig 6 10 Fig 6 3 and Tab 6 7 Fig 6 9 Cable gland Fig 6 10 Terminals for the connection of the power supply and the outputs Tab 6 6 Connection of the power supply AC DC terminal connection terminal conn...

Page 35: ...2 4 Uext 4 26 4 V Uext 0 021 A Rext 4 V example Uext 12 V Rext 0 380 current during transmitter error Ifault 0 mA HART mode I1 1 I1 2 Uext 10 24 V current during transmitter error Ifault 0 mA passive...

Page 36: ...tor G800 G800L G800P G800LP B1 B4 6 8 10 12 B1 B4 5 7 9 11 Uext 5 24 V Rc k Uext Ic mA Ic 1 4 mA binary output Reed relay G800 G800L G800P G800LP B3 B4 9 11 B3 B4 10 12 Umax 48 V Imax 100 mA RS485 G80...

Page 37: ...see chapter 18 Recommended configuration of the alarm output If the transmitter is disconnected from the power supply the binary output indicates an alarm state Tab 7 Circuit of the current output Ex...

Page 38: ...chapter 14 6 9 Sensor Module SENSPROM The sensor module contains important transducer data for the operation of the transmitter with the transducers It is con nected to the connector strips above the...

Page 39: ...tened 7 3 Installation 7 3 1 Wall Installation Fix the instrument mounting plate 2 to the wall with the 4 screws 4 see Fig 7 1 Fix the transmitter to the instrument mounting plate 2 with the 2 screws...

Page 40: ...ducer measuring channel B transducer measuring channel A power supply outputs AV AVS ARS AR BVS BRS BR BV PE N L L1 L 1 2 3 4 6 7 8 5 9 10 11 12 14 15 13 FLUXUS G801 A10 AA G801 A10 AP AV AVS ARS AR B...

Page 41: ...ransducer cable and brush it back Insert the end of the extension cable into the housing Screw the gasket ring side of the basic part into the housing Fix the cable gland by screwing the cap nut onto...

Page 42: ...2 Remove the left blind plug for the connection of extension cable see Fig 7 6 Open the cable gland of the extension cable The compression part remains in the cap nut Push the extension cable through...

Page 43: ...rents can flow to the transmitter via the extension cables Remove the left blind plug for the connection of extension cable see Fig 7 7 Open the cable gland of the extension cable The compression part...

Page 44: ...he terminals of the transmitter see Fig 7 7 and Tab 7 4 Fig 7 7 Connection of the extension and transducer cable to the junction box JB01 Attention The external shield of the extension cable must not...

Page 45: ...6 Tab 7 5 Terminal assignment transducer cable terminal connection V transducer core VS transducer shield RS transducer shield R transducer core Attention Observe the Safety Instructions for the Use i...

Page 46: ...e power supply AC DC terminal connection terminal connection PE earth PE earth N neutral L DC L1 Phase 100 240 V AC L DC Attention Observe the Safety Instructions for the Use in Explosive Atmosphere s...

Page 47: ...1 k DIN EN 60947 5 6 NAMUR binary output open collector G801 A10 FF B1 6 B1 5 Uext 5 30 V Rc k Uext Ic mA Ic 2 100 mA Ioff 0 8 mA binary output open collector G801 A10 A G801 A10 P B1 B4 6 8 10 12 B1...

Page 48: ...10 A G801 A10 P 14 A 13 B 120 termination resistor shield 15 Tab 8 Circuit of the current output Ex ib output transmitter external circuit remark internal circuit connection passive current loop Ex ib...

Page 49: ...serial interface of the PC If the RS232 cable cannot be connected to the PC use the RS232 USB adapter The RS232 adapter the RS232 cable and the RS232 USB adapter are part of the serial data kit optio...

Page 50: ...If transducers are replaced or added the sensor module also has to be replaced or added Insert the sensor module into the socket of the measuring channel to which new transducers have to be connected...

Page 51: ...distance between the inner edges of the transducers see section 3 3 and Fig 8 1 For the determination of the flow direction see section 11 8 Select the installation instructions that correspond to th...

Page 52: ...in reflection arrangement only one transducer mounting fixture has to be mounted see Tab 8 1 Fig 8 4 Scope of delivery Note When measuring in diagonal arrangement with 2 beams and displaced X arrangem...

Page 53: ...to length pipe circumference at least 120 mm Ratchet clasp Cut the tension strap to length pipe circumference at least 120 mm Insert approx 100 mm of the tension strap into parts 1 and 2 of the clasp...

Page 54: ...pe and insert it into the clasp see Fig 8 12 Tighten the tension strap Tighten the screw of the clasp Quick release clasp Insert the tension strap into the tension strap clamp and the metal spring see...

Page 55: ...pipe see Fig 8 9 On a horizontal pipe mount the tension strap clamp on the side of the pipe if possible Mount the metal spring on the opposite side of the tension strap clamp if necessary Place the te...

Page 56: ...upplied clasp Band clamp clasp Insert the tension strap into tension strap clamp 1 see Fig 8 19 Place the tension strap around the pipe and insert it into the clasp see Fig 8 20 Tighten the tension st...

Page 57: ...pe diameter 80 mm or on pipes that are not subjected to significant temperature fluctuations Position the clasp the metal spring if necessary and tension strap clamp 1 on the pipe Mount the metal spri...

Page 58: ...8 22 Tension strap with the metal spring and the tension strap clamp Fig 8 23 Ratchet clasp with tension strap Fig 8 24 Ratchet clasp with tension strap Note To release the screw and the tension stra...

Page 59: ...rain see Fig 8 27 Put coupling foil or some coupling compound for a short term installation on the contact surface of the transducers The coupling foil can be fixed to the contact surface with a small...

Page 60: ...cer mounting fixture is described transducers in reflection arrangement Overview of the mounting steps step 1 disassemble the transducer mounting fixture Variofix C step 2 mount the tension straps wit...

Page 61: ...n of the rail without a clasp Cut the tension strap to length pipe circumference at least 120 mm Insert approx 100 mm of the tension strap into one of the slots of the tension strap clamp and bend it...

Page 62: ...bend it Bend both ends of the tension strap see Fig 8 36 b Repeat the steps for the second tension strap Position the tension strap at the distance s see Fig 8 36 Put the rail on the tension strap cla...

Page 63: ...cant temperature fluctuations Place the tension strap around the pipe see Fig 8 40 Position the metal spring if mounted the ratchet clasp and the tension strap clamp On a horizontal pipe mount the ten...

Page 64: ...rews Fig 8 40 Tension strap with the metal spring the ratchet clasp and the tension strap clamp on the pipe Fig 8 41 Ratchet clasp with tension strap Fig 8 42 Ratchet clasp with tension strap Note To...

Page 65: ...tensioning screws slightly The end of the screw has to be placed above the hole in the transducer see Fig 8 44 Correct the transducer distance if necessary see section 11 6 1 and section 11 6 2 Tight...

Page 66: ...C as follows Use a lever tool to remove the cover Insert the lever tool in one of the four openings of the cover see Fig 8 47 Press the lever tool against the fixture Bend the cover outwards and relea...

Page 67: ...r crosswise They are installed on the entire pipe circumfer ence Pipe damping mats can be installed to reduce the propagation of acoustic noise in the pipe wall if it is not possible to keep the minim...

Page 68: ...emove the protective foil bit by bit and fix the damping mat to the pipe at the same time Use a roller to fix the damping mat to the pipe Press the roller on the damping mat First move the roller from...

Page 69: ...ping mats see Tab 9 1 Tab 9 1 Installation length of the transducer and pipe damping mats Diagonal arrangement Reflection arrangement lS1 installation length of the transducer damping mat diagonal arr...

Page 70: ...ngth is 1010 mm Transducer damping mats are installed on the pipe lengthwise The pipe damping mats can be installed on the pipe lengthwise or crosswise In the example they are installed lengthwise Num...

Page 71: ...cer damping mat reflection arrangement lR installation length of the pipe damping mat lS1 length of the transducer mounting fixture 2 20 mm lS2 2 length of the transducer mounting fixture 2 20 mm lR l...

Page 72: ...e they are installed crosswise Number of transducer damping mats The transducer damping mats are installed along the pipe circumference with a width of 2 x 1400 mm The number of transducer damping mat...

Page 73: ...soon as the transmitter is connected to the power supply the serial number of the trans mitter is displayed for a short time It is not possible to enter any data while the serial number is displayed...

Page 74: ...a detailed overview of the menu structure see annex A SYSTEM settings contains the following menu items set clock libraries dialogs and menus measurement outputs storing signal snap network serial tr...

Page 75: ...scroll and select a list item in the upper line Press key to scroll and select a value for the selected list item in the lower line Press ENTER function HotCode see section deactivation language sele...

Page 76: ...of the Power Supply As soon as the measurement begins all current measuring parameters will be stored in a non volatile INIT resistant EPROM The measurement will be interrupted if the power supply fai...

Page 77: ...ructions for the Use in Explosive Atmosphere see document SIFLUXUS Note During the parameter input the transducers have to be connected to the transmitter Note The parameters will only be stored when...

Page 78: ...temperature will be displayed directly see section 11 2 5 Enter the sound speed of the pipe material Press ENTER Note Enter the sound speed of the material i e longitudinal or transversal speed which...

Page 79: ...and therefore for the calculation of the recommend ed transducer distance as well as for the interpolation of the compressibility coefficient of the gas at the beginning of the measurement During the...

Page 80: ...dard transducer parameters stored in the transmitter Select Special Version to enter the transducer parameters The transducer parame ters have to be provided by the manufacturer Press ENTER Note If st...

Page 81: ...he accuracy of the entered pipe and fluid parameters If the displayed transducer distance is adjusted press ENTER The measuring for the positioning of the transducers is started The amplitude of the r...

Page 82: ...d in the program branch Parameter and the transducer dis tance has been optimized afterwards as described in section 11 6 2 it is recommended to write down the sound speed for the next measurement By...

Page 83: ...ction of the values to be displayed and for the setting of the output options see chapter 12 For more measuring functions see chapter 13 11 8 Detection of the Flow Direction The flow direction in the...

Page 84: ...the units of measurement can be displayed with a prefix during the measurement The unit of measurement of the operating volumet ric flow rate is displayed with an A the unit of measurement of the stan...

Page 85: ...be activated after an initialization of the transmitter All Channels The measured values of all channels measuring and calculation channels are displayed The next active channel is se lected after mi...

Page 86: ...re shown in the status line The quality and precision of the ongoing mea surement can be estimated display explanation Mass Flow designation of the physical quantity A 8 879 m3 values of the totalizer...

Page 87: ...ameter tem perature dependence pressure dependence ok is equal to the expected value 20 of the expected value 20 of the expected value unknown cannot be measured R flow profile information about the f...

Page 88: ...be determined There are two flow totalizers one for the positive flow direction one for the negative flow direction The unit of measure ment used for totalization corresponds to the volume or mass uni...

Page 89: ...reset to zero manually Press key to scroll through the upper line and display the totalizers The totalizer value will be displayed in the upper line here the volume which has passed through the pipe...

Page 90: ...user are selected two values will have to be entered If Cut off Flow absolut and user is selected only one value will have to be entered Enter 0 zero to switch off the checking for outliers Enter a l...

Page 91: ...normal to display and output the flow velocity with profile correction Select uncorr to display the flow velocity without profile correction Press ENTER If uncorr is selected it has to be confirmed e...

Page 92: ...ted The damping factor has to be set separately for each of the two measuring channels Two cut off flow values for each calculation channel can be defined The cut off flow is not based on the flow vel...

Page 93: ...nnel will be invalid This message will be displayed after the parameterization of the calculation channel if the calculation function n is selected The measured values of the calculation channel here...

Page 94: ...channels are activated or deactivated in the same way as the measuring channels Press ENTER If a measuring channel that is needed for an activated calculation channel is not activated a warning will...

Page 95: ...an intervention in the measurement a command or key BRK 13 11 1 Defining a Program Code A program code will remain valid as long as no other valid program code is entered or the program code is not de...

Page 96: ...entered program code is not valid an error message will be displayed for a few sec onds If the entered program code is valid the measurement will be stopped To execute a command it is sufficient to e...

Page 97: ...will be signalized acoustically This signal can be de activated see section 14 1 3 in Acoustic Signal 14 1 1 Activation Deactivation of the Data Logger 14 1 2 Setting the Storage Rate The storage rate...

Page 98: ...ct the moment at which the storing has to start Promptly The storing starts immediately On full 5 min The storing starts in the next full 5 minutes On full 10 min The storing starts in the next full 1...

Page 99: ...er memory can be dis played If on is selected and the data logger is activated the amplitude of the measured signal will be stored together with the measured values Press ENTER If on is selected and t...

Page 100: ...serial number of the transmitter is displayed in the upper line The available data logger memory will be displayed in the lower line here 18 327 addition al measured values can be stored Press key BRK...

Page 101: ...elect the character to be used to separate columns semicolon or tabulator Press EN TER Enter HotCode 232 0 see section 10 4 Set the transmission parameters in the 4 scroll lists Press ENTER baud baud...

Page 102: ...rminal Program Start the terminal program Enter the transmission parameters into the terminal program see section 14 2 4 The transmission parameters of the terminal program and the transmitter have to...

Page 103: ...art the program FluxData V3 0 or higher on the PC Select the menu Options Serial interface Select the serial interface used from the PC e g COM1 Click Protocol Click OK Enter the transmission paramete...

Page 104: ...be stored Click OK Select the path on which the data should be stored and enter the file name Click Save The file will be stored with the file extension flx Example DEVICE GX80X XXXXXXXX MODE ONLINE...

Page 105: ...s cannot be activated for the physical quantity flow velocity these columns will not be generated Offline Transmission of Data During the offline transmission of data columns will only be created if a...

Page 106: ...rts for the following material data material parameters transversal and longitudinal sound speed typical roughness fluid parameters min and max sound speed kinematic viscosity density compressibility...

Page 107: ...Select Special Funct Install Material or Install Medium Press ENTER This error message will be displayed if the coefficient memory does not contain an area for user defined materials fluids Partition...

Page 108: ...eed of the fluid Press ENTER This display will only be indicated if user is selected Enter the kinematic viscosity of the fluid Press ENTER Enter the density of the fluid Press ENTER Tab 15 2 Material...

Page 109: ...ces Select Special Funct SYSTEM settings Libraries Extended Library Press ENTER Select on to activate the extended library Press ENTER Select Special Funct Install Material or Install Medium Press ENT...

Page 110: ...no Press ENTER Change the values if necessary 15 4 Deleting a User Defined Material Fluids To delete a user defined material fluid proceed as follows Select Special Funct Install Material or Install...

Page 111: ...list can be edited see section 15 5 1 15 5 3 Select End of Edit to stop editing Press ENTER Select yes to store all changes of the scroll list or no to quit the menu item without storing Press ENTER...

Page 112: ...d from the scroll list Press ENTER All materials fluids of the current scroll list will be displayed in the lower line Select the material fluid Press ENTER The material fluid will be removed from the...

Page 113: ...ress ENTER The new date is displayed Select ok to confirm the date or new to set the date again Press ENTER Select Special Funct SYSTEM settings Dialogs Menus Press ENTER Note The settings will be sto...

Page 114: ...be requested in the program branch Parameter If off is selected 1 bar will be used for all calculations Note For documentation purposes it is useful to enter the fluid pressure even if the transmitter...

Page 115: ...will only be indicated if lb ft has not been selected as the unit of measurement for the density Select mm s or cSt as the unit of measurement for the kinematic viscosity Press ENTER Select m s or fps...

Page 116: ...profile correction Press ENTER For further information see section 13 6 A lower limit for the flow velocity can be entered see section 13 5 An upper limit for the flow velocity can be entered see sec...

Page 117: ...h the following keys increases the contrast reduces the contrast Note After an initialization of the transmitter the display is reset to medium contrast Select Special Funct Instrum Inform to display...

Page 118: ...perUser mode it is possible to define some flow parameters profile bounds correction of the flow velocity for the specific application or measuring point Attention The SuperUser mode is intended for e...

Page 119: ...item SYSTEM settings For the definition of the correction of the flow velocity see section 17 3 2 Example profile bound for the laminar flow 1 500 profile bound for the turbulent flow 2 500 At Reynold...

Page 120: ...ed If the transmitter is switched off without starting a measurement the entered correction data will be lost Attention The correction of the flow velocity is still active after the deactivation of th...

Page 121: ...the totalizer is reduced with the number of decimal places For setting the behavior of the totalizers after reaching the max value see section 13 3 1 17 8 Manual Reset of the Totalizers If the manual...

Page 122: ...play During the Measurement In the SuperUser mode the following parameters can be displayed during the measurement besides the normal informa tion see section 12 3 Select on to activate the display of...

Page 123: ...be indicated if the output has not been installed yet Select yes Press EN TER If the output has already been installed select no to reconfigure it or yes to uninstall the output and to return to the p...

Page 124: ...t item or other range to enter the output range manually If other range is selected enter the values Output MIN and Output MAX Press ENTER after each input This error message will be displayed if the...

Page 125: ...measured during the time interval t0 t1 see Fig 18 1 The error value will be transmitted Fig 18 1 Error output Tab 18 3 Examples for the error output list item for the error output output signal error...

Page 126: ...within the output range Press ENTER If the multimeter displays the entered value the output functions correctly Select yes to repeat the test no to return to SYSTEM settings Press ENTER Select Reed R...

Page 127: ...of the source item will be displayed Zero Scale Val is the measured value corresponding to the lower limit of the output range as defined in section 18 1 1 Enter the highest expected measured value F...

Page 128: ...item can be totalized a request will be in dicated whether the frequency output is to be configured as a pulse output Select yes Press ENTER Enter the upper limit of the frequency Press ENTER The low...

Page 129: ...selected as the physical quantity The use of the pulse output is not possible in this case because integrating the flow velocity does not result in a reasonable value Enter the pulse value The unit of...

Page 130: ...ing function NO Cont The alarm is energized if the switching condition is true and de energized if idle NC Cont The alarm is de energized if the switching condition is true and energized if idle Note...

Page 131: ...measured values This might cause an apparent switching delay when the measured value changes marginally less than 2 decimal places In this case the switching accuracy of the output is higher than the...

Page 132: ...cond to avoid humming i e fluctua tion of the measured values around the value of the switching condition An alarm output of the type NON HOLD will be activated if the switching condition is met It wi...

Page 133: ...g Select the menu item SHOW RELAIS STAT Select on to activate the alarm state indica tion The alarm state indication is structured as follows Rx with x being the number of the alarm output and a picto...

Page 134: ...Track down the situations when the error is displayed Contact FLEXIM The transmitter does not react when key BRK is pressed during the measurement A program code has been defined Press key CLR and ent...

Page 135: ...sit formation in the pipe Avoid measuring points in the vicinity of deformations and defects on the pipe and in the vicinity of welds Measure the temperature at the measuring point and make sure that...

Page 136: ...ther measuring point to check whether the results are better Because pipes are never rotationally symmetric the flow profile is affected Change the transducer position according to the pipe deformatio...

Page 137: ...600 6553 5 m s This display will only be indicated if Other Material is selected selection whether the pipe is lined selection of the lining material This display will only be indicated if Lining yes...

Page 138: ...This display will only be indicated if Other Medium is selected or no data set for the selected fluid is stored in the transmitter input of the fluid temperature default 20 C input of the fluid pressu...

Page 139: ...ned selection of the physical quantity selection of the unit of measurement for the physical quantity input of the duration over which a floating average of the measured values has to be determined ra...

Page 140: ...ly be indicated if Pulse Output is activated input of the pulse width range 1 1000 ms This display will only be indicated if Pulse Output is activated Alarm output Activation of an Alarm Output This d...

Page 141: ...uid scroll lists SYSTEM settings Libraries Material list selection of the displays for the arrangement of the material scroll list pipe and lining materials SYSTEM settings Libraries Medium list selec...

Page 142: ...in the program branch Measuring user Only the entered transducer distance will be displayed if the recom mended and the entered transducer distances are identical auto Only the recommended transducer...

Page 143: ...the default limit of 2 5 cm s will be used user input of a limit x input of the cut off flow for positive measured values range 0 12 7 cm s 0 127 m s default 2 5 cm s 0 025 m s This display will only...

Page 144: ...s activated x activation of the display of the totalizer sum This display will only be indicated if the SuperUser mode is activated x activation of the display of the last valid measured value This di...

Page 145: ...the displays for the input of the standard conditions for the gas measurement This display will only be indicated if Special Funct SYSTEM set tings Measuring Gas Measuring is activated input of the p...

Page 146: ...n of the decimal marker for floating point numbers x selection of the character for column separation x selection of the serial interface default RS232 This display will only be indicated if the trans...

Page 147: ...ains measured values and the transmitter is connected to a PC via a serial cable error message that there is a problem with the serial transmission display of the data transmission progress Delete Mea...

Page 148: ...ed material input of the constants for the transversal sound speed of the material The number of constants depends on the function selected above input of the constants for the longitudinal sound spee...

Page 149: ...dependence of the fluid parameters selection of a user defined fluid selection whether the user defined fluid is to be edited or deleted This display will only be indicated if the selected fluid alrea...

Page 150: ...he gas program code selection of the displays for the input of a program code defining a program code input of the break code program code input of the access code the first three digits of the progra...

Page 151: ...e bar g bar relative psi a pound per square inch absolute psi g pound per square inch relative Density unit of measurement description g cm gram per cubic centimeter kg cm kilogram per cubic centimete...

Page 152: ...r m hl s hectoliter per second hl or m Ml d Megalit d megaliter per day Ml or m bbl d barrel per day bbl bbl h barrel per hour bbl bbl m barrel per minute bbl USgpd US gal d gallon per day gal USgph U...

Page 153: ...r minute klb 1 lb 453 59237 g 1 t 1000 kg heat flow rate Heat quantity totalized unit of measurement description unit of measurement W Watt Wh or J kW kilowatt kWh or kJ MW megawatt MWh or MJ GW gigaw...

Page 154: ...0 D N 5 0 0 D N 6 0 0 D N 8 0 0 D N 1 0 0 0 1 m 3 h 1 0 1 0 0 1 0 3 1 0 4 l m i n 1 0 0 1 0 1 0 3 1 0 4 1 0 5 0 1 1 l s 1 0 1 0 0 1 0 3 1 1 0 g a l m i n 1 0 0 1 0 3 1 0 4 1 0 5 1 0 0 b b l d 1 0 3 1...

Page 155: ...D N 4 0 0 D N 6 0 0 1 0 0 b b l d 1 0 3 1 0 4 1 0 5 1 0 6 1 1 0 g a l m i n 1 0 0 1 0 3 1 0 4 1 0 l m i n 1 0 0 1 0 3 1 0 4 1 0 5 1 1 0 m 3 h 1 0 0 1 0 3 1 0 4 0 1 1 l s 1 0 1 0 0 1 0 3 S t r m u n g...

Page 156: ...ong m s cflow Carbon Steel steel normal 3 230 5 930 trans Stainless Steel steel stainless 3 100 5 790 trans DUPLEX duplex steel 3 272 5 720 trans Ductile Iron ductile iron 2 650 trans Asbestos Cement...

Page 157: ...1 5 encrusted 1 5 3 Fluid display Explanation Sound speed m s Kinematic viscosity mm s Density g cm acetone acetone 1 190 0 4 0 7300 Ammonia NH3 ammonia NH3 1 386 0 2 0 6130 Gasoline gasoline 1 295 0...

Page 158: ...R134 Freon 522 0 2 1 2400 R22 Freon R22 Freon 558 0 1 1 2130 Crudeoil hi API crude oil light 1 163 14 0 0 8130 Crudeoil low API crude oil heavy 1 370 639 5 0 9220 30 H2SO4 sulphuric acid 30 1 526 1 4...

Page 159: ...583 481 54 0 595270779 0 9677784 0 80 68 928 411 41 0 184177693 0 819764 10 64 534 422 6 0 19880993 0 8446627 20 60 824 433 08 0 213649217 0 8656106 30 57 632 442 93 0 228709745 0 883441 40 54 841 452...

Page 160: ...Fig D 1 FLUXUS G801 see Fig D 2 Flameproof enclosure If the O ring is defective or the housing does not seal anymore change it see Tab D 1 Housing with increased safety If the O ring is defective or t...

Page 161: ...x UMFLUXUS_G8V4 6 1EN 2018 10 10 161 Tab D 1 transmitter O ring item number FLUXUS G800 upper housing gasket Adalet 120 mm 990735 1 FLUXUS G800 lower housing gasket Adalet 95 mm 990735 2 FLUXUS G801 O...