Auto-tuning (PID
Parameter Calibration)
(continued)
The auto-tuning (calibration) procedures are as follows:
(1) Press the CONFIG Button and select the auto-tuning mode [CALi]
shown below by turning the Control Knob:
(2) Press the Control Knob and select YES by turning the Control Knob:
(3) Press the Control Knob again to start the auto-tuning:
Restoring the Default
Settings
To change the current settings back to the default settings, do as
follows:
(1) Press the CONFIG Button and select the default setting [dEFA]
shown below by turning the Control Knob:
(2) Press the Control Knob and confirm the change to the default
settings by using the Control Knob as shown below:
(3) Terminate the configuration mode as described in Enter/Terminate
Configuration Mode.