4-12
Section 4
Operation
Temperature Offsetting
(continued)
(4) Terminate the configuration mode as describe in Enter/Terminate
Configuration Mode.
Maximum Heating Rate
Setting
The higher the heating rate, the faster the heat up time but the wider
the temperature overshoot and undershoot. Therefore, if you want to
reduce the temperature fluctuation, you need to limit the heating rate to
a certain degree by limiting maximum heating rate.
• When the heat up time is too slow, check whether the maximum
heating rate is set too low.
If, for example, the current heating rate is 100% but you want to set the
heating rate limit at 60%, do as follows:
(1) Press the CONFIG Button and select the heating rate limit mode by
turning the Control Knob:
• The [LiMt] display for the maximum heating rate setting shown
above and the [LiSt] display for the configuration checking mode
shown in 4-9 look similar but should not be confused.
(2) Press the Control Knob and check the current heating rate limit: