4-13
Section 4
Operation
Maximum Heating Rate
Setting (continued)
(3) Change the limit to the desired value by turning the Control Knob.
(4) Save the desired limit by using the Control Knob as shown below:
(5) Terminate the configuration mode as described in Enter/Terminate
Configuration Mode.
Change the High/Low
Temperature Limits
Note that the default setting of the low temperature limit [t-L] is 0°C and
the high temperature limit [t-h] 350°C. If needed, however, you can set
your own temperature limits. If, for example, you want to set the high
temperature limit at 200°C, do as follows:
(1) Press the CONFIG Button and select the temperature limit mode
shown below by turning the Control Knob:
(2) Press the Control Knob and check the current high temperature
limit:
(3) Change the limit to the desired value by turning the Control Knob.