![Fisher Scientific Isotemp RT Operation Manual Download Page 28](http://html1.mh-extra.com/html/fisher-scientific/isotemp-rt/isotemp-rt_operation-manual_2296591028.webp)
4-11
Section 4
Operation
Temperature Offsetting
The temperature shown on the Actual Temperature Display is
measured by a temperature sensor inside the unit. However, this
temperature can be different from the temperature of your own
thermometer which you may use as a standard for your specific
applications. If needed, you can offset such temperature differences at
0.1°C interval.
• The allowed range of the temperature offsetting is -10 ~ +50C.
• See Technical Specification for the temperature difference depend-
ing on the top plate temperature.
If, for example, the actual temperature of the top plate is 100°C but the
displayed temperature is 95°C, you can match the displayed
temperature with the actual temperature of the unit by selecting the
offset value of +5°C and save it as described below:
(1) Press the CONFIG Button and select the offsetting mode by turning
the Control Knob as shown below:
(2) Press the Control Knob and select the offset value by turning the
Control Knob:
(3) Save the desired offsetting value by pressing the Control Knob as
shown below: