4-3
Section 4
Operation
Temperature Setting
(t mode) (continued)
(3) Push the Control Knob to save the desired temperature setting. The
following displays will appear in sequence:
• You can change the temperature setting even during operation.
• You can cancel the temperature setting any time before pushing the
control knob.
• Changing the temperature setting is allowed only within the low and
high temperature limits. If changing the temperature setting cannot
be done properly, check the low and high limits first. See Changing
the High and Low Temperature Limits.
• You can set the low and high temperature limits at any value bet-
ween 0°C and 350 °C.
Temperature Setting
(r mode)
The temperature setting procedures under heating rate priority mode (r
mode) are as follows:
(1) Select the desired heating rate by turning the Heater Knob:
(2) Select the desired temperature as described in ‘r mode’ above.
• You can change the heating rate even during operation.
• Note that wide temperature fluctuation (overshoot and undershoot)
may happen if the heating rate is high.
• You can adjust the heating rate ranging from 0 to 100% at 1% inter-
val.