Temperature Control
Mode Selection
(continued)
(1) Press the CONFIG Button and select the temperature control mode
selection [CoEF] shown below by turning the Control Knob:
(2) Press the Control Knob and check the current temperature control
mode:
(3) Select the Slow Mode by turning the Control Knob.
(4) Save the desired mode by using the Control Knob as shown below:
(5) Terminate the configuration mode as describe in Enter/Terminate
Configuration Mode.
Auto-tuning (PID
Parameter Calibration)
The PID parameters for temperature control can be automatically tuned
to take your specific operating circumstances into consideration.
• Before starting the auto-tuning process, make sure to place the load
you want to use on the top plate.
• To cancel the auto-tuning during the process, turn off the main pow-
er switch. If cancelled, the Optimal Mode will be selected
automatically.
• Note that the auto-tuning process takes approximately 40 to 60
minutes.
• If you want to control the temperature using calibrated parameters,
select the user mode [uSEr] under temperature control mode [CoEF]
described previously.