. Sleep Data
Turn the SmartDial to the Sleep Data setting and press to enter.
Alternatively, press and hold the SmartDial for
7
seconds. Turn the
SmartDial to either the Centers for Medicare and Medicaid (CMS) Report
(US specific) or Patient Report
and press to enter.
Turn the SmartDial to scroll through each of the following statistics:
• Total Compliant Hours
.
• Average Compliant Hours over the last
30
nights
. Press the
SmartDial to access the average for the last
7
nights and again for the
last night.
• Percentage of nights used greater than
4
hours over the last
30
nights
. Press the SmartDial to access the percentage for the last
7
nights.
• The number of nights elapsed since therapy commenced
.
• Check Sum
.
• AHI for last
30
,
7
, and
1
nights.
• Leak for last
30
,
7
, and
1
nights.
•
90
th
percentile pressure for last
30
,
7
, and
1
nights.
• SensAwake for last
30
,
7
, and
1
nights.
• Exit the Sleep Data setting.
NOTE: A symbol will appear on the Home Screen when the sleep data
is within the compliance requirements. Once CMS requirements have
been met, the data within the CMS sub-menu will stop logging.
• Exit the Sleep Data setting.
12
12
. SensAwake On/Off
The SensAwake setting is available on both Auto and Premo models,
but may be restricted by your healthcare provider. To adjust, turn the
SmartDial to the SensAwake setting
and press to enter. Turn the
SmartDial to select On or Off and press again to accept the change.
NOTE: This setting is not available with the Novo model.
13
13
. User Preferences
Turn the SmartDial to the User Preferences setting and press to enter.