8. TROUBLESHOOTING

If you feel that your device is not operating correctly, please refer to the following suggestions. If the 
problem persists, please consult your healthcare provider. Do not attempt to repair the device yourself.

Problem

Possible Cause

Solution

I can not turn the 
pressure on or off

The device is likely to be in the Advance 
Mode of the Menu System and the selection 
may not be at the 

12

 o’clock position.

Turn the selection circle to the 

12

 o’clock 

position and press the SmartDial to turn the 
pressure On or Off.

Insufficient air is delivered 
from the device

The Chamber Lid may not be correctly fitted 
to the device. Ensure the Water Chamber is 
in the device.

Refer to Section 

1

, Part 

1

.

“Error” is displayed on 

the Home Screen of the 
device

An error may have been detected with the 
device.

Disconnect from power, then connect and 
re-start device. If the error persists, record the 
number displayed and contact your healthcare 
provider for further instructions.

AlarmTunes does not 
work

The Alarm may be set to “off”.
The Alarm may be set to “buzzer”.

F&P Studio software may not have been 
installed.
The Alarm volume may not be set loud 
enough.
The InfoUSB may not be inserted all the way 
into the USB port.

Refer to Section 

1

, Part 

2

 for instructions on 

the Alarm Setting.
Once F&P Studio software has been installed 
onto your computer and your favorite songs 
selected, download onto your InfoUSB and 
then ensure the InfoUSB is inserted into the 
USB Port of the device. 

The Heater-Plate and/or 
Water Chamber base is 
warm to the touch even 
though the device isn’t 
being used

The power supply is located directly 
underneath the Heater-Plate and in stand-by 
mode generates approximately 

5

 W of power, 

which causes the feeling of warmth. The 
Water Chamber is also fully insulated by the 
device which can cause heat to be retained. 
This is perfectly normal and should not cause 
concern.

If concerned about overheating, unplug the 
device from the power supply when it is not 
being used.

I think my Water Chamber 
is leaking because there 
is a buildup of water on 
the Heater-Plate

After therapy the device will cool, which may 
result in condensation forming inside the 
Water Chamber and pooling on the Heater 
Plate.

Before each use remove the Water Chamber 
and dry the basin of the device with a cloth. 
If the water buildup becomes excessive, 
please contact your healthcare provider. 

I don’t think my 
Humidifier is working

Check the Humidity setting is above 

1

. Check 

your Water Chamber as it may be empty.

Refer to Section 

1

, Part 

2

 for more 

information on the humidity controls. 

The InfoUSB symbol 

 

is constantly flashing on 
the Home Screen

The InfoUSB may not be inserted correctly.

Take the InfoUSB completely out of the USB 
Port, then insert again until it clicks.

The device seems noisy; 
there is a hissing noise 
coming from the device

The Chamber Lid may not be properly fitted, 
causing air to leak.

Remove the Chamber Lid and replace again 
so that it is completely flush with the device, 
by following the instructions set out in 
Section 

1

Summary of Contents for ICON

Page 1: ...Use and Care Manual ...

Page 2: ... F P ICON Menu System A 4 Simple Mode 1 On Off and Ramp A 4 2 Humidity A 4 3 Sleep Data A 4 Advanced Mode 4 On Off and Ramp A 5 5 Humidity A 5 6 Alarm On Off A 5 7 Alarm Time A 6 8 Alarm Volume A 6 9 Clock Time A 6 10 Brightness A 6 11 Sleep Data A 7 12 SensAwake On Off A 7 13 User Preferences A 7 13i Flight Mode A 8 13ii Display Pressure A 8 13iii Altitude Adjustment A 8 3 Communicating Sleep Dat...

Page 3: ...led water up to the maximum water line which is marked on the side of the Water Chamber CAUTION Do not use the device without a Water Chamber Do not use hot water and water should be replaced before each use IMPORTANT The Water Chamber cannot be disassembled To replace the Water Chamber the arrow on the circular chamber grip should face toward the Elbow Fit the Chamber Lid by lining up the large d...

Page 4: ...rt Breathing Tube come together with the electrical connectors on the Elbow Please refer to Oxygen Elbow user instructions if attaching an Oxygen Elbow From the Home Screen a quick press to the SmartDial will start therapy The Home Screen will light up and the clock face of the Menu System will rotate to indicate therapy is starting If Ramp is required press the SmartDial again and hold for 3 seco...

Page 5: ... setting is Humidity level 4 If using without water or where low power consumption is required e g for aircraft and trucking use set the humidity level to 0 3 3 Sleep Data Press and hold the SmartDial for 7 seconds to access the Sleep Data setting Turn the SmartDial to scroll through each of the following statistics Total Compliant Hours Average Compliant Hours over the last 30 nights Press the Sm...

Page 6: ...he humidity level cannot be adjusted if Flight Mode is set to on The Boost control offers three additional levels 1 2 3 to optimize the Humidity setting Press and hold the SmartDial for 3 seconds while in the Humidity setting and the Boost symbols will appear Turn to select low 1 medium 2 or high 3 and press again to accept the change CAUTION The Boost control should only be used in extreme situat...

Page 7: ...to accept the change To adjust the click volume of the SmartDial press and hold for 3 seconds while in the Alarm Volume setting The Volume symbol will be present turn to select the desired volume and press again to accept the change 9 9 Clock Time Turn the SmartDial to the Clock Time setting and press to enter Turn the SmartDial to select the hour minutes 12 hr or 24 hr time After each selection p...

Page 8: ...apsed since therapy commenced Check Sum AHI for last 30 7 and 1 nights Leak for last 30 7 and 1 nights 90th percentile pressure for last 30 7 and 1 nights SensAwake for last 30 7 and 1 nights Exit the Sleep Data setting NOTE A symbol will appear on the Home Screen when the sleep data is within the compliance requirements Once CMS requirements have been met the data within the CMS sub menu will sto...

Page 9: ...me Screen Time is the default setting however this can be changed to show pressure cmH2O or hPa if required Turn the SmartDial to the Display Pressure setting and press to enter Turn to select On or Off and press again to accept the change If a different unit of measure is required press the SmartDial and hold for 3 seconds Turn the SmartDial to select either cmH2O or hPa and press again to accept...

Page 10: ... USB Port of a computer connected to the Internet The first time you insert the InfoUSB 2 into your computer open Windows Explorer and navigate to the drive called FPHCARE 1 To install an InfoUSB detector on your computer open this folder and double click on the Setup exe file Follow the on screen instructions Upon successful installation of the InfoUSB 2 application and upon detection of an InfoU...

Page 11: ...er Empty the vinegar solution and rinse the Water Chamber thoroughly with water Disconnect the Elbow from the device and the Outlet Seal can be easily removed by lightly pushing it from the back of the device into the Water Chamber housing The Elbow and the Outlet Seal can be cleaned and washed with mild dishwashing detergent then rinsed and dried Please place the Outlet Seal before the Elbow in s...

Page 12: ...ith respiratory failure or chronic obstructive pulmonary disease The safety and effectiveness of the auto adjusting device has not been established in patients with congestive heart failure obesity hypoventilation syndrome or central sleep apnea 3 1 2 Adverse effects Nosebleeds ear and sinus discomfort may occur from the use of positive pressure therapy 3 2 Warnings 3 2 1 To avoid electric shock f...

Page 13: ...ternatively consult your healthcare provider Do not block the flow through the device as this may cause the motor to overheat Do not block the exhaust flow holes of the mask as they are designed to allow a continuous flow of air out of the mask and if blocked CO2 re breathing may occur In the event of power failure machine malfunction or if the device is turned off remove the mask immediately as t...

Page 14: ... water damage to the device Remove the Water Chamber from the device before filling Do not fill the Water Chamber above the maximum level Do not move carry transport or store the device with water in the Water Chamber 3 3 2 General Only use the device within the Operating Conditions specified in Section 2 Part 4 Position the device so the power cord connection to the power supply is easily accessi...

Page 15: ...4 5 Dynamic Pressure Stability Breath Rate BPM Test Pressure Pmin 4 0 cmH2 O Pmin 1 4 Pmax Pmin 8 0 cmH2 O Pmin 1 2 Pmax Pmin 12 0 cmH2 O Pmin 3 4 Pmax Pmin 16 0 cmH2 O Pmin 20 0 cmH2 O Novo Dynamic Pressure Stability cmH2O 10 0 76 0 84 0 90 0 99 1 04 15 0 76 0 84 0 91 0 99 1 04 20 1 16 1 22 1 28 1 38 1 44 Auto Premo Dynamic Pressure Stability cmH2O 10 0 13 0 16 0 22 0 28 0 33 15 0 27 0 26 0 31 0 ...

Page 16: ...00 F 38 C NOISE LEVEL Sound Pressure Level 29 dBA Average Sound Power Level 37 dBA WATER CHAMBER VOLUME 420 mL up to the maximum fill line STANDARDS COMPLIANCE IEC60601 1 1988 A1 A2 AS NZS 3200 1 0 1998 EN60601 1 1990 A1 A2 UL60601 1 2003 DATA RECORDING The InfoUSB records up to 365 days of summary efficacy data 7 days of detailed efficacy data AHI Leak Pressure and Leak data 50 Hz as well as cumu...

Page 17: ... to activate ThermoSmart Technology Part may vary depending on country 6 DEVICE AND CONSUMABLES DISPOSAL INSTRUCTIONS Device Disposal Instructions This device contains electronics Please do not discard as regular rubbish Dispose of electronics according to local guidelines Consumable Disposal Instructions Place the mask Breathing Tube and Water Chamber in a rubbish bag at the end of use and discar...

Page 18: ...n ensure the InfoUSB is inserted into the USB Port of the device The Heater Plate and or Water Chamber base is warm to the touch even though the device isn t being used The power supply is located directly underneath the Heater Plate and in stand by mode generates approximately 5 W of power which causes the feeling of warmth The Water Chamber is also fully insulated by the device which can cause h...

Page 19: ...any Tel 49 7181 98599 0 Fax 49 7181 98599 66 India Tel 91 80 4284 4000 Fax 91 80 4123 6044 Irish Republic Tel 1800 409 011 Italy Tel 39 06 7839 2939 Fax 39 06 7814 7709 Japan Tel 81 3 3661 7205 Fax 81 3 3661 7206 Northern Ireland Tel 0800 132 189 Spain Tel 34 902 013 346 Fax 34 902 013 379 Sweden Tel 46 8 564 76 680 Fax 46 8 36 63 10 Switzerland Tel 0800 83 47 63 Fax 0800 83 47 54 Taiwan Tel 886 2...

Reviews: