To avoid choking, or inhalation of a foreign object:
• Never place any non-approved objects into any opening of the device, Breathing Tube or mask.
• Ensure the recommended filter is fitted to the device before use.
• Ensure the Breathing Tube is positioned so it can not become entangled with the body or furniture
during sleep.
3
.
2
.
3
To avoid burns:
• Do not fill the Water Chamber with hot water.
• After use, wait for the water to cool before touching, carrying or emptying the Water Chamber.
• Do not touch the heater-plate shortly after the device has been operating.
• Do not touch the water in the Water Chamber while the device is operating.
• Position the ThermoSmart Breathing Tube so it is uncovered and free from bedding or other
materials. Do not lie on it and avoid prolonged skin contact.
3
.
2
.
4
Other:
• Ensure the device is stored and operated on a surface, level or below head height, to prevent
water entering the tubing and the device enclosure.
• The device is not intended to be used as a life-supporting device.
• The device complies with the electromagnetic compatibility requirements of
IEC60601-1-2
. In
certain circumstances the device may affect or be affected by nearby portable mobile radio
frequency communication equipment, due to the effects of electromagnetic interference. If this
should happen, try moving your device or the location of the equipment causing interference, or
alternatively consult your healthcare provider.
• Do not block the flow through the device as this may cause the motor to overheat.
• Do not block the exhaust flow holes of the mask as they are designed to allow a continuous flow
of air out of the mask and if blocked CO
2
re-breathing may occur.
• In the event of power failure, machine malfunction or if the device is turned off, remove the mask
immediately as the flow through the mask may be insufficient to clear all exhaled gas, which may
result in CO
2
re-breathing occuring that can be hazardous.
• The device is not suitable for use in environments with flammable anaesthetic mixtures with air or
oxygen or nitrous oxide.
3
.
2
.
5
To ensure optimal therapy:
• Do not operate the device, Water Chamber or Breathing Tube if it is dropped, damaged or not
working as intended.
• Pressure adjustments should only be made by a qualified healthcare provider.
• Only use masks, breathing tubes and accessories compliant with ISO
17510
-
2
, distributed for use
with this device, and recommended by Fisher & Paykel Healthcare or your healthcare provider.
• It is recommended to only use the parts that have been specified in this manual.
• On models without automatic Altitude Adjustment, ensure the altitude level is manually adjusted to
ensure optimal pressure delivery.
• Do not use the device without a Water Chamber in place.
• If using the device without water in the Water Chamber it is recommended to set the humidity level
to zero.
• It is recommended to use distilled water to extend the life of your Water Chamber. This is likely to
prevent residue buildup on the chamber base.