- OPTION OUTPUT, CAMERA I, CAMERA II
Sockets Output 12V DC,
max. 12W/outlet, max. 24W
collective load
- ILLUMINATION CONTROL
3 pin Socket
Control Voltage Cold Light
- TABLE AIR OUT
Table Air Outlet
- HEAD VACUUM OUT
Arm Vacuum Outlet
- PIPETTE
Pipette Connector
- COMPRESSED AIR
Compressed Air Input
- INPUT VACUUM (option)
Connector ext. Vacuum
- Mains Voltage
230V, 240V / either 50Hz or 60 Hz *
100V, 120V / either 50Hz or 60 Hz *
(* The Placer Control Box is factory-
provided pre-adjusted internally for a
specific mains voltage in combination
with a specific frequency)
- Mains Fuse
230V, 240V: 1,25A slow
100V, 120V: 2,5A slow
- Power consumption
max. 200VA
- Vacuum generation
typical. -0,4bar/ -5,8psi
Alternatives:
- Vacuum generation (compressed air):
6 bar (-0,5/ +1 bar) free of water, oil
and particles
- External vacuum supply (option):
General function range between -0,4
and -0,8bar **,
(**factory-provided pre-adjusted
internally for standard vacuum-
switch actuation point at -0,6bar, in
case of lower grad vacuum the
vacuum switch must be readjusted)
- TABLE AIR OUT
1bar/ 14,5psi (internal pump)
- Weight:
Placer Control Box 9,3 kg, Placer Arm
max. 1kg
- Dimension:
285 x 290 x 145mm (l x w x h)
- Length Placer-Arm-Connector:
ca. 1,8
Environment:
- Operating conditions:
15-35
°
C/ 59-95
°
F ambient
temperature
20-75% relative humidity