Instructions for Set-Up
4.1
Initial Set-Up of the FINEPLACER
®
4.1.1
Preparation of the Basic Components and Subassemblies
1.
See safety instructions, chapters 2.5 to 2.9.!
2.
Set-up should take place in a low-dust room. The base plate (70) on which the
positioning table (29) floats, and the under side of the positioning table (29)
should be cleaned thoroughly with a clean, dry, soft cloth before use. Put the
positioning table onto the base plate (70). Keep the base plate clean at all
times!
3.
When the equipment is set up, level the ground plate by adjusting the leveling
screws (15) below the base plate. After interconnecting cables and hoses to
the Placer Control Box (39), verify that the level has a minimum lateral
descent so that the positioning table will not float uncontrolled in any
direction.
4.
The alignment of the superimposed image of the component as well as the
subject may be impaired by any light from the background. A dark wall or
screen behind the erected pivot arm (27) will improve this. Take care not to
blind operators in adjacent work stations.
Differences in Vision Modules:
For FINEPLACER
®
lambda, different
objective lenses are available. E.g.
in case of insufficient image, first
compare the S/N of the Finetech
with the delivery note entry.
4.1.2
Preparation of the Positioning Table's Height Adjustable Tracks
•
A fundamental rule is that the height of the substrate's placement surface
must be the same as that of the pivot arm axis.
•
The Tracks (13 and 23) of the positioning table (29) are parallel and contain
V-shaped slots for receiving substrates, substrate support plates or heating
plates.
TABLE 3
FINEPLACER® lambda
field diameter
x
y
A6
min
0,55mm
0,45mm
max
6,7mm
5,4mm
A7
min
0,18mm
0,14mm
max
ø 1.25 mm
A7+
min
0,1 mm
0,08 mm
max
ø 0.6 mm