24. System & Error Codes
Code
Description
Err 1
Time value error
Err 2
Date value error
Err 3
Exceed manual zero
Err 4
Offset out of range/unstable during power on (5 minutes for OIML and NTEP mode)
Err 5
Load cell not detected
Err 6
Tare operation error
Err 7
Logic error. Hi limit is lower than Lo limit (and Hi is not = 0)
Err 8
Logic error. Lo limit is higher than Hi limit (and Hi is not = 0)
Err 13
Exceed maximum power on zero
Err 19
Capacity or division setting error (Division set is higher than 10000d)
Err 22
Manual Zero and Tare stability error
Err 23
Capacity setting error, Capacity 1 > Capacity 2
Err 24
Division setting error, e1 > e2
Err 25
Span gain is too low
Err 26
Not able to obtain stable status for longer than 10 sec
Err 27
•
The calculated value per e of Cal 2 varies more than 1% as of Cal 1. Properly a
load cell problem.
•
The mass value of Cal 2 is less than 150% of Cal 1
Err 28
The maximum accumulation limit is exceeded
Err 29
Standard deviation data exceed memory size (300 transactions).
Err 30 ~ 33
WIFI command execution error
Err 34
Cannot communicate with WIFI module
Err 35, 36
WIFI command execution error
Err 38, 39
WIFI command execution error
Err 42
Cannot communicate Bluetooth module
Err 43
Bluetooth module execution command error
Err 44
RS485 Alibi report output setting error
Err 46
Customized Operation PLU number does not exist
Err 60
USB Disk error or not exist
Err 61
USB Disk cannot find or create a file
Err 62
USB Disk cannot find or create folder path
Err 63