30
9.8 Tare Functions
9.8.1 Manual & automatic tare
This instrument is equipped with the below various tare functions. These functions can be selected in F12.
a.
Off: - Manual Tare
b.
Auto: - Instrument assumes the first stable weight (≥ 20d or 20d
1
) applied is a container and tare off its
weight. When the container is removed and the gross weight is zero, tare effect will be cancelled.
c.
Contin (continuous auto tare): - All stable weight (≥ 20d or 20d
1
) applied will be tared off. When all loads
are removed and the gross weight is returned to zero, the tare effect will be cancelled.
Notes: -
•
If Contin is selected, instrument will prompt for delay time (0.0 ~ 9.9 second
.
Default = 1.5 second).
•
Delay time is the time duration from when a stable weight is detected until it is tared off.
•
Input the preferred delay time and press
[Print/M+]
to save.
9.8.2 Repetitive/multiple tare
This instrument is equipped with multiple tare operation capability. Settings: -
•
Set F13 = On to allow multiple tare operation. If it is the case, instrument will permit multiple tare
operation, and tare effect can only be cancelled when there is no load .
•
Set F13 = Off to disable multiple tare operation. If this is the case, the tare effect can only be canceled
when the container is removed, and the gross weight is zero.
9.8.3 Preset tare
Preset tare allows a pre-determined tare weight value can be inputted via numeric keys.
9.8.3.1 Introduce a preset tare value
Input the pre-determined tare weight value and press
[Print/M+]
to enter. After the pre-determined tare value
has been inputted: -
•
Instrument displays the preset tare value inputted.
•
Net indicator
appears to indicate the value being displayed is net weight.
9.8.3.2 Cancel a preset tare value
A preset tare value can be canceled by either one of the below methods: -
•
Remove all loads and press
[Tare]
.
•
Input a zero preset tare value and press
[Print/M+]
.
9.9 Select the Preferred Weight Unit
7
This instrument supports kg, g, and lb. The desired weight units should be enabled in F9 before use. Contact
your dealer if in doubt.
Disregarding the setting of
F9
, weight unit gram (g) is available only when 3 or 4 decimal places (0.000 or
0.0000) is selected in F81. Contact your dealer for more details.
To select the preferred weight unit, press
[Unit]
until the preferred weight unit appears.
The weight unit employed before powering off will be employed again when powered on again.
7