56
Appendix B: - Customized PC Output Protocol
This instrument supports customized PC output protocol. Under this mode: -
•
2 data separation types
•
7 control codes, and
•
26 different transaction data
are available from instrument.
B.1 Customized PC Output Protocol Setting Procedures
1.
At Weighing mode, go to F16 and select the preferred UART number depends on the output comport number.
2.
Select the preferred transmission type.
3.
Select PC, then set baud rate, parity, and data length.
4.
Select Custom in Protocol page. Press
[Print/M+]
.
5.
Instrument displays one of the data separators.
6.
Press
[Unit]
to shift between data separator Comma (comma) and SemiCo (semicolon) and press
[Print/M+]
.
Notes
: -
•
Data separator is a symbol inserted between transaction data and used by a computer program to separate various data.
•
No data separator is inserted between in front or after control commands.
•
Separator in front and after a control code (C1 ~ C6) will not be surpassed. For example: -
STXdata
,
data
,
dataETX
•
Exception of above is when control codes are in between 2 data. For example: -
data
,
CR LF data
7.
Instrument displays item number, then followed by the content page.
Note: -
Item number means the output sequence, e.g., Item 1 = the first content to output, Item 3 = the third content to output.
8.
Press
[Unit]
or
[MR]
until the preferred content appears and press
[Print/M+]
. Refer to
Customized PC Output Content Table
for more information.
9.
Repeat steps
7
and
8
to include other transaction data or control.
10.
To complete and save a customized output, select
End
and press
[Print/M+]
.
11.
Then select the output time interval. Refer to paragraphs
8.1 & 8.2
for more information.
12.
Press
[Print/M+]
to save.