background image

 

Sweep 

Press the 

SWEEP

 key on the front panel to enable the scanning function, and 

fy6200  can  output  the  scanning  waveform  from  ch1  channel.

 

The  process  of 

gradient scanning for a specific parameter object at a specified time from the initial 
value to the termination value.

 

For sine wave, square wave, sawtooth wave and 

arbitrary wave, scanning output can be generated. 

 

2-2 Sweep parameter setting interface 

Sweep Object 

   

The  FY6200  can  output  sweep  waveforms  from  the  CH1  channel.  The 

parameters  that  can  be  scanned  are:  frequency,  amplitude,  offset,  duty  cycle. 
Switching object types can be switched via the 

OBJE

 soft key. 

 

 

In the frequency scanning mode, the signal generator changes the output from 
the starting frequency to the ending frequency within the specified sweep time. 

 

In the amplitude scanning mode, the signal generator changes the output from 
the starting amplitude to the ending amplitude within the specified scanning 
time. 

 

In the offset scanning mode, the signal generator changes the output from the 
start offset to the end offset within the specified scanning time. 

 

In duty cycle scanning mode, the signal generator changes the output from 
the start duty cycle to the end duty cycle within the specified scanning time. 

Summary of Contents for FY6200 Series

Page 1: ...FeelElec User s Manual FY6200 Series Fully Numerical Control Dual Channel Function Arbitrary Waveform Generator Rev2 0 April 2019...

Page 2: ......

Page 3: ...ous publications FeelElec shall not be liable for any errors that may be contained in this manual or for any incidental or consequential loss resulting from the information and deductive functions pro...

Page 4: ...litude 18 Set Offset 19 Set Duty Cycle Square 20 Set Phase 21 Enable Channel Output 22 Example Output sine waveform 23 Generate burst waveform Digital Modulation Frequency meter counter 30 Enable freq...

Page 5: ...completely independent two way DDS signal and four way TTL level output which can generate 31 preset waveform signals such as sine wave square wave triangle wave sawtooth wave pulse wave white noise a...

Page 6: ...ase difference can be precisely adjusted Standard channel tracking function When tracking is turned on all parameters of the two channels can be updated according to the user s configuration at the sa...

Page 7: ...There has Manual Trigger internal CH2 Trigger and External Trigger for your options It can output 1 1048575 pulse trains Various modulation types AM FM PM ASK FSK and PSK modulations Storage feature i...

Page 8: ...ing container for damage Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical...

Page 9: ...f the current function Please refer to the User Interface section for details 2 Manu Buttons Corresponding to the menu displayed on the left press this soft key to activate the corresponding menu 3 Fu...

Page 10: ...witching Support prompt tone off on Supports multi machine cascading Supports master slave switchover in cascaded state Supports dual channel power on default output state setting 4 Direction key When...

Page 11: ...put of the CH1 channel and can be switched to the CH1 parameter setting interface in any interface Press this button the CH2 light will turn on and the CH2 output will turn on At this point the CH2 co...

Page 12: ...serial driver Through the upper computer software or user defined programming Warning In order to avoid damaging the instrument the voltage range of the input signal of the ext in terminal shall not...

Page 13: ...the whole machine is less than 3W Power On Turn on the power switch after the power cord is connected The generator will execute self inspection The LCD will show welcome interface after the inspectio...

Page 14: ...urrent channel selected for operation 2 Current waveform type selected Displays the name of the currently selected feature For example ch1 amplitude modulation waveform means that the currently select...

Page 15: ...ding FREQ button to highlight the amplitude display value and change the parameter through the direction key and knob 8 Offset Displays the DC offset of the currently selected channel waveform Press t...

Page 16: ...Dimensions...

Page 17: ...ct Output Channel CH1 CH2 keys are used to switch ch1 or CH2 to the currently selected channel When power on ch1 is selected by default and displayed in yellow on the top half of the screen Press the...

Page 18: ...Wave Rectification User defined waveform Press WAVE to change waveform selected Or rotate ADJ Knob under waveform switching status to change waveform The waveform diagram displays on the screen Pressi...

Page 19: ...fications The factory default setting is 10KHz Press the FREQ soft key to highlight the frequency parameter At this time use the direction key and knob to set the parameter value use the direction key...

Page 20: ...he knob to modify the value Key Points 1 What is the difference between the amplitude in Vpp and the value in Vrms Answer Vpp is the unit of signal peak value Vrms is the unit of signal effective valu...

Page 21: ...key to move the cursor to select the position to be edited and then rotate the knob to modify the value The minimum adjustment accuracy of offset is 1mV 0 001V When the output frequency is lower than...

Page 22: ...djusting range of duty cycle of this instrument is 0 01 99 99 b Press the OK key in the duty cycle adjustment state and the duty cycle will be initialized to 50 0 2 Press duty cycle repeatedly to swit...

Page 23: ...0 to 359 99 and the phase resolution is 0 01 The default value is 0 1 The starting phase of the screen display is the default value or the previously set phase 2 Press the PHAS soft key to highlight...

Page 24: ...generator works in the waveform parameter setting menu and the currently selected channel is CH1 Pressing the CH1 button will switch the output between off on 2 The signal generator works in other fun...

Page 25: ...0 020 000 000 000kHz 4 Set the Amplitude Press AMPL to highlight the amplitude value Press the direction key and rotate the adjustment knob to obtain the following data AMPL 02 500V 5 Set the offset v...

Page 26: ......

Page 27: ...al signal or manual pulse signal as modulation wave to modulate signal It can realize FSK ASK PSK digital modulation trigger pulse series output and AM FM PM analog signal modulation functions The mod...

Page 28: ...switch the modulation modes including FSK frequency keying ASK amplitude keying PSK phase keying Trigger controllable burst output AM Amplitude Modulation FM Frequency modulation PM Phase modulation E...

Page 29: ...ternal AC access external signals in the form of AC coupling through the EXT IN port Manual click OK key manually as modulation signal External DC access external signals in the form of DC coupling th...

Page 30: ...can be adjusted In the AM mode you can adjust the modulation rate of the amplitude modulation 0 to 200 In FM mode frequency offset parameters can be adjusted In the PM mode the degree of phase deviat...

Page 31: ...by corresponding soft keys Press COUNTER soft key to set the number of pulse strings When the number of pulses is selected use the direction key and parameter adjustment knob to change the number of s...

Page 32: ...real time The minimum frequency of measurement is 0 01Hz gate time 100s Press COUNTER soft key to enter the external pulse counting function At this time COUNTER soft key will be changed to FREQ Press...

Page 33: ...Gate Time Press the GATE soft key to select the gate time of the measurement system The default is 1s When measuring low frequency signal 10s or 100s gate time can be selected Strobe time Frequency r...

Page 34: ...re frequency amplitude offset duty cycle Switching object types can be switched via the OBJE soft key In the frequency scanning mode the signal generator changes the output from the starting frequency...

Page 35: ...highlight the start parameter press the direction key and rotate the adjustment knob to adjust to the start value specified by the user as shown in the following example START 10 000V Offset scanning...

Page 36: ...ng example END 20 000V Offset scanning press the END soft key to highlight the END parameter press the direction key and rotate the adjustment knob to adjust to the cut off value specified by the user...

Page 37: ...start value scanning time 100 Logarithmic scanning In the logarithmic scanning mode the parameters of the output signal of the instrument change in logarithmic mode Take the frequency sweep as an exam...

Page 38: ...ncy to high frequency When the starting frequency the ending frequency the signal generator scans from high frequency to low frequency When the starting frequency the ending frequency the signal gener...

Page 39: ...ut Operation method Press the VCO button to enter the function interface After the scan object start cutoff and scan mode settings are completed connect the external scan signal to the Counter port on...

Page 40: ...ronization configuration etc SAVE The current waveform parameters can be stored in 20 groups of storage areas of the system LOAD The preset parameter information in the system can be loaded into the c...

Page 41: ...m waveform parameter information into the current system status Select the SAVE soft key on the right to store the current system parameters in the corresponding storage location Select the right LOAD...

Page 42: ...ace press the right parameter soft key to set the system s working mode Press the soft key and select the system prompt language as Chinese Press the Eng soft key and select the system prompt language...

Page 43: ...the FY6200 include waveform frequency amplitude offset duty cycle and can be set separately WAVE The soft key is highlighted and the waveform type of CH2 changes synchronously with the waveform type o...

Page 44: ...restart the instrument 4 If you are still unable to use the product properly please contact FeelElec Signal generator CH2 locked 1 Check whether the signal generator works in synchronous state Press t...

Page 45: ...nd can reach 1uhz so as to ensure the adjustment accuracy of the instrument at high frequency For example the instrument can output 10 000000000001MHz high precision signal Frequency Accuracy 20ppm Fr...

Page 46: ...short circuit condition DC Offset Offset Range Frequency 20MHz 10V Frequency 20MHz 2 5V Offset Resolution 1mV Minimum Phase Feature Adjustable Range 0 359 99 Minimum resolution 0 01 TTL Output TTL Le...

Page 47: ...uency VCF Voltage control range VCA Voltage controlled bias voltage controlled duty ratio VCO Feature Amplitude modulation AM or frequency modulation FM can be performed by an external analog signal M...

Page 48: ...y Display Type 3 2 inch TFT color LCD display Storage and Load Quantity 20 sets Location 01 to 20 Power on by default to enter 01 storage location parameters Connector Interface mode USB to serial int...

Page 49: ...l be exempt from any claims that may arise from this to the company s managers and employees affiliated agents distributors etc including various costs compensation fees attorney fees etc Appendix C a...

Reviews: