Example: output sine wave
This section mainly introduces how to output a sine wave from the [ch1]
connector (frequency 20KHz, amplitude 2.5vpp, offset 1.6vdc, initial phase 90.9 °)
1. Select output channel:
Press CH1 to select CH1. Now all characters and
border of the channel is displayed in yellow.
2. Select the Sine:
Press
WAVE
button to select Sine. Then the diagram of
Sine displays on the screen.
3. Set the frequency:
Press
FREQ
button to highlight the frequency value.
press the direction key to adjust the cursor to the "2" position as shown in the
figure below, and rotate the adjustment knob to obtain the following data.
FREQ
:
00’0
2
0.000’000’000kHz
4. Set the Amplitude
: Press
AMPL
to highlight the amplitude value, Press the
direction key and rotate the adjustment knob to obtain the following data.
AMPL
:
02.500V
5.
Set the offset voltage
:
Press
OFFS
to highlight the offset value.
Press the
direction key and rotate the adjustment knob to obtain the following data.
OFFS
:
01.600V
6.
Set Phase
: Press the
PHAS
soft key to highlight the "phase" parameter. Press
the direction key and rotate the adjustment knob to obtain the following data.
PHAS
:
090.90°
7. Enable channel output:
Press the
CH1
key to turn on the ch1 output light,
and the [CH1] connector outputs the sine wave signal in the current configuration.
8. Observe the output waveform
: Connect the [CH1] of fy6200 to the
oscilloscope with BNC connecting wire. The following figure shows the waveform
observed by the oscilloscope.