- 148 -
INF-TN1FSH-E
Bi-directional range:
Forward and reverse ranges configurable independently
Hysteresis: 0 to 20% of working range configurable
Working range applicable to digital output
Auto-2 ranges:
Forward 2 ranges configurable independently
Hysteresis: 0 to 20% of working range configurable
Working range applicable to digital output
Flow switch:
Lower and upper switching points configurable independently
Acting point applicable to digital output
Total switch:
+total switching point configurable
Acting point applicable to digital output
Physical specifications
Enclosure protection:
Flow Transmitter: IP67,
Detector: IP67
Mounting:
Flow Transmitter: wall mount
Detector: Clamped on pipe surface
Acoustic coupler:
Silicon compound (RTV)
Material:
Flow Transmitter: aluminum alloy
Detector: PBT for housing, aluminum alloy for frame and SS for
fastening belt
Sensor cable (FLY6):
RF coaxial cable (double shielded)
External sheath: Black flame-resistant vinyl
External diameter: About 7.3 mm
Terminal treatment: Water-resistant BNC connector (detector side),
M3.5 amplifier terminal (Flow Transmitter side)
Mass: About 90 g/m
Temperature sensor cable (FLY7):
3-core shield cable
External sheath: Gray flame-resistant vinyl
External diameter: About 6.9 mm
Terminal treatment: Round waterproof connector (detector side),
M3.5 amplifier terminal (Flow Transmitter side)
Mass: About 56 g/m
Dimensions:
Flow Transmitter: H240
×
W247
×
D134 mm
Detector:
H70
×
W57
×
L360 mm (FSWS12)
H72
×
W57
×
L540 mm (FSWS21)
H90
×
W85
×
L640 mm (FSWS40)
H82
×
W71
×
L258 mm (FSWS50)
Mass:
Flow Transmitter: 5 kg
Detector: 1.7 kg (FSWSl2), 1.9 kg (FSWS2l), 5 kg (FSWS40),
1.5 kg (FSWS50)
Loader software for PCs
Equipped as standard:
• Works on PC/AT compatible machines.
• Operation on PC98-series machines (NEC) cannot be guaranteed.
• Operation on self-made PCs or shop-brand PCs cannot be
guaranteed.
• Major functions: Setting/changing of various parameters for the
main unit
If no flow velocity profile output is selected, the
following functions are not available.
“Detailed setting” and ”flow velocity profile display”
in pulse Doppler measurement
“Detailed setting” and ”receved signal display” in
Transit time measurement
• O/S: Windows 2000/XP
• Memory requirement: 128MB or more
• Disk unit: Windows 2000/XP-compatible CD-ROM drive
• Hard disk drive capacity: Free space of 52MB or more
Note: PC loader communication cable (type ZZP*TK4H6253,
Specifications: D-sub 9 pin receptacle, cable length 3m) is
separately required.
Detector frame installation fixture
Installation fixture is provided to facilitate the positioning of the frame to
the piping.
Select a desired type from the following according to the detector to be
used.
Type Applicable
detector
ZZP*TK7M7071C1 FSWS12
ZZP*TK7M7071C2 FSWS21
ZZP*TK7M7071C3 FSWS40
Note: The installation fixture cannot be used for detector type FSWS50,
which is not provided with a frame.
Maximum measurement range in hybrid mode
When stainless steel is selected as pipe material, nominal wall thickness is Sch20s, and the fluid is water
<Maximum measurable flow velocity>
Unit: m/s
<Maximum measurable flow rate>
Unit: m
3
/h
Diameter FSWS12 FSWS21 FSWS40 FSWS50
FSWS12 FSWS21 FSWS40 FSWS50
40A
6.56
33.6
50A
6.52
52.7
65A
5.31
72.1
80A
4.65
86.5
90A
4.12
102
100A
3.69
7.25
118
231
125A
3.08
6.08
147
289
150A
2.63
5.20
179
354
200A
2.04
4.05
7.77
239
474
908
250A
3.30
6.38
604
1168
300A
2.78
5.41
735
1428
350A
2.51
4.90
820
1598
400A
2.20
4.31
951
1858
450A
3.84
2118
500A
3.48
3.48
2358
2358
550A
3.17
2618
600A
2.91
2879
650A
2.71
3096
700A
2.52
3357
750A
2.35
3618
800A
2.21
3879
850A
2.08
4140
900A
1.97
4400
1000A
1.77
4902