- 12 -
INF-TN1FSH-E
2.2.2. Mounting position
The instrument can be mounted horizontally or in any other position. However, pay attention to the following.
(1) Allow fluid to fill the pipe and keep it flowing at all times.
Pump
Air tends to pile up.
Good
Good
The pipe
may not be
filled with fluid.
The pipe may not be filled with fluid.
(2) Mount the flowmeter within
±
45
°
from the center plane in the case of horizontal pipe run.
Mount it at an arbitrary position on the outer periphery in the case of vertical pipe run.
Horizon
Pipe
45
45
(3) Avoid mounting the flowmeter in a position where the pipe is deformed, or on a flange or welded part.
Includes welded part.
Does not include welded part.
Welded part
Welded part
Overlaid on welded part.
Not overlaid on welded part.
×
○
○