INF-TN1FSH-E
- 65 -
4.4.9.3. Total preset
Description
Total preset value can be set as follows.
Reset the total mode, and the preset value is restored.
•
F: Total preset: Total preset value in forward direction
•
R: Total preset: Total preset value in reverse direction
(Setting range: 0 to 9999999999.999)
Note: Set the total preset value in a state where the total mode is set to stop.
Operation
(example)
When setting forward direction to 1000m
3
and reverse direction to 2000m
3
Key operation
Description
Display
FUNC TOTAL
Display “TOTAL MODE”
TOTAL MODE
TOTAL STOP
"
or
#
Select “F: TOTAL PRESET”
F: TOTAL PRESET
0.000 m
3
ENTER 1 0 0 0 ENTER
Enter numeric value enter mode, enter
“1000” using ten keys, and press ENTER.
F: TOTAL PRESET
1000.000 m
3
#
Select “R: TOTAL PRESET”
R: TOTAL PRESET
0.000 m
3
ENTER 2 0 0 0 ENTER
Enter numeric value enter mode, enter
“2000” using ten keys, and press ENTER.
R: TOTAL PRESET
2000.000 m
3
ESC
Return to “TOTAL MODE”
TOTAL MODE
TOTAL STOP
ENTER
"
or
#
ENTER
Enter select/enter mode, select “TOTAL
RESET,” and press ENTER.
TOTAL MODE
TOTAL RESET
ESC
Display the measurement, reflecting the
setting.
(Measurement display screen)
Preset value
Total value
Time
Reset
Reset
Reset