![FCI ST100 series Installation, Operation & Maintanance Manual Download Page 9](http://html.mh-extra.com/html/fci/st100-series/st100-series_installation-operation-and-maintanance-manual_547358009.webp)
ST100 Series Flow Meter
GENERAL
Fluid Components International LLC
3
Instrument
Measuring Capability
ST1XX Models:
Flow rate, total flow and temperature
STP1XX Models:
Flow rate, total flow, temperature and pressure
Basic Style
ST100:
Insertion, single-point
ST100L:
In-line (spool piece), single-point
ST102:
Insertion, dual-element system
ST110:
Insertion, single-point with VeriCal™ capability
ST112:
Insertion, dual-element system with VeriCal capability
STP100:
Insertion, single-point with pressure measurement
STP102:
Insertion, dual-element system with pressure measure-
ment
STP110:
Insertion, single-point with pressure measurement and
VeriCal capability
STP112:
Insertion, dual-element system with pressure measure-
ment and VeriCal capability
Flow Measurement Range
Insertion Style:
0.25 SFPS to 1000 SFPS [0,07 NMPS to 305 NMPS]
ST100L In-line:
0.0062 SCFM to 1850 SCFM
[0.01 Nm3/h to 3,140 Nm3/h]
– Air at standard conditions; 70 °F and 14.7 psia [21 °C and 1,01325
bar (a)]
Temperature Measurement Range
Up to 850 °F [454 °C] commensurate with element; see Operating
Temperature in Flow Element specification
Pressure Measurement Range (STP Models)
Available Ranges:
0 psig to 50 psig [0 bar (g) to 3,4 bar (g)]
0 psig to 160 psig [0 bar (g) to 11 bar (g)]
0 psig to 500 psig [0 bar (g) to 34 bar (g)
0 psig to 1000 psig 0 bar (g) to 70 bar (g)]
Environmental Conditions
Maximum Relative Humidity: 93%
Maximum Elevation: 6561’ [2000m]
Media:
All gases that are compatible with the flow element material
Accuracy:
Flow:
Gas Specific Calibration: ± 0.75% reading, ± 0.5% full scale
SpectraCal Gas Equivalency: Typically ± 4% reading, ± 0.5% full
scale;
gas conditions specific to application will determine accuracy; utilize
FCI’s online tool, AVAL, to evaluate your application and provide
expected accuracy
Temperature:
± 2 °F [± 1.1 °C] (display only, flow rate must be
greater than 5 SFPS [1,5 m/sec])
Pressure (STP Models):
± 0.25% full scale pressure range
Temperature Coefficient
With optional temperature compensation; valid from 10% to 100%
of full scale calibration
Flow:
Maximum ± 0.015% of reading / °F up to 850 °F
[± 0.03% of reading / °C up to 454 °C]
Repeatability
Flow:
± 0.5% reading
Temperature: ± 1 °F [± 0.6 °C] (flow rate must be greater than 5 SFPS
[1,5 NMPS])
Turndown Ratio
Normally factory set and field adjustable from 2:1 to 100:1
within calibrated flow range; up to 1000:1 possible with factory
evaluation of application
Temperature Compensation
Standard:
± 30 °F [± 16 °C]
Optional:
± 100 °F [± 55 °C]
Agency Approvals
FM, FMc: Class I, Division 1, Division 2, Hazardous Locations;
Groups B,C,D,E,F,G
ATEX and IECEx (See page 87)
CPA, NEPSI
Calibration
Performed on NIST traceable flow stands and equipment
Storage Temperature
-76 to 150°F [-60 to 65 °C]
Flow Element
Material of Construction
All-welded 316L stainless steel; Hastelloy-C optional
Operating Pressure
Metal ferrule:
1000 psig [69 bar (g)]
Teflon ferrule:
150 psig [10 bar (g)] (200 °F [93 °C] maximum)
Fixed Connection NPT:
1000 psig [69 bar (g)]
Fixed Connection Flanged:
per flange rating
Operating Temperature (Process)
ST100, ST102 Insertion Style
All Flow Elements (– FPC, – FP and – S):
-40 °F to 350 °F [-40 °C to 177 °C]
-40 °F to 500 °F [-40 °C to 260 °C]
-40 °F to 850 °F [-40 °C to 454 °C]
ST110, ST112 Insertion Style
– FP Style Flow Element:
-40 °F to 350 °F [-40 °C to 177 °C]
-40 °F to 500 °F [-40 °C to 260 °C]
STP Series Insertion Style
All Flow Elements (– FPC, – FP and – S):
-40 °F to 257 °F [-40 °C to 125 °C]
Technical Specifications
STP Models
See pressure sensor specifications
Same as ST type above or per pressure sensor specification, whichever is lower
Model ST100L
In-line style
Tubing
Sch 40 Pipe
Sch 80 Pipe
Psig
Bar(g)
Psig
Bar(g)
Psig
Bar(g)
F Style Sensor
1
″
2400
165
2500
172
3000
207
1 1/2
″
1750
121
2500 *
172 *
2
″
1500
103
2250 *
155 *
S Style Sensor
1
″
2400
165
2500
172
2500
172
1 1/2
″
1750
121
2500 *
172 *
2
″
1500
103
2250 *
155 *
*
1 1/2
″
and 2
″
Sch 80 available by special order only; contact FCI
3/4
″
pipe also available by special order