OPERATION
ST100 Series Flow Meter
26
Fluid Components International LLC
The Internal Delta-R Resistor Check may be programmed ahead of time using the ST100 configuration software application. The user may run
and save a Test Log (to the ST100 internal FRAM) automatically at these user-defined intervals:
Day of Month (1-28)
Every Nth Day (1-255)
Day of Week (0=Sun)
Every Day
To run the check immediately on demand click
Run Check Now
.
IDR files are viewed in the “Test Logs” tab. These files may be added to the microSD card for further analysis by selecting the “Add to SD
Card Logs.” The SD Card must be manually removed to transfer these IDR log files to a PC. Use a card reader as described in "Remove mi-
croSD Card from the Instrument and Manually Transfer Files with a Card Reader" on page 24.
The IDR log file is always named “DLTRLOG” and is amended every time a new scheduled test is initiated. Process data log files are always a
new file with a unique file name as shown.
Internal Delta-R Log File Results Example (Data Formatted and Titles Added using Microsoft Excel)
IDR Log and Process Data Log Files on microSD Card Example