12.
Close the pressure release valve.
13.
Turn PUMP switch to ON and fill the pressure vessel with oil. Tighten the
small plug when bubble free oil begins to run out the side.
14.
Adjust AIR TO PUMP to adjust pump pressure as desired.
15.
Loosen the small top plug located on top of the magnetic drive housing to
bleed any air bubbles located in the magnetic drive system. (If the
consistency reading is very high when you start the test it may be because
there is entrapped air in the mag drive). Then close the plug shut.
16.
Place the motor assembly onto the slurry cup plug and tighten the
associated set-screws to prevent it from being dislodged or wobble.
17.
Press START TEST
18.
Activate AUTO SHUTDOWN and select desired shutdown value.
Be careful not to over-pressure the cylinder. As the heat
increases during a test so does the pressure. Pressure is not
released automatically and requires that the pressure be
monitored and adjusted by a technician.
7.2 Stopping a Thickening Time Test
The steps listed below are for experienced users who are familiar
with consistometer operation.
1.
Press STOP TEST using the touch screen or allow AUTO SHUTDOWN to
activate.
2.
Activate the cooling button if not done so automatically.
3.
Undo coupling tube set screws if installed and remove the motor and set it
down away from any interference.