Wiring should be done by a qualified installer in accordance with
local electrical codes.
The instrument should be securely connected to a separate earth
ground. The ground wire must be larger in diameter than the
supply conductors.
An 8BC or larger fire extinguisher to fight electrical and oil fires
should be placed within 50 feet of the consistometer.
Some components such as touch screen LCD monitors may be removed from the
instrument prior to shipment and shipped in a separate container to prevent damage.
This device must be reinstalled before operating the instrument. The cables on the
touch screen monitor must be plugged into the correct location on the supplied
control box. These connections include a 15-pin SVGA cable, USB, and AC (120-
240VAC) power. The touchscreen power should be powered by the instrument and
not into a separate power source prevent electrical ground loops.
If a printer is included with the instrument, it may be connected to the rear of the
control box. Connect the USB connector on the printer cable to an available USB
port on the control box. The printer must also be connected to a suitable power
source. A USB flash drive may also be connected to the instrument in lieu of a
printer. Test data may be downloaded to the USB drive and then transferred to a PC
for archival storage.
The control box must be properly wired to the instrument for
operation.
Connect the 5-pin round power plug into the outlet labeled COMPUTER on the
back of the instrument. In addition, a USB type-A to Type-B cord must be
connected from the back of the control box to the back of the instrument, also
labeled COMPUTER.