Model 415C & 415SGS Instruction Manual
D01669102
April 2019, Revision A
12
An uninterruptable power supply (UPS) is recommended for any
areas experiencing poor power conditions. If you are installing a
UPS, connect the output of the UPS to the outlet labeled UPS on
your instrument. If not, you MUST CONNECT A POWER JUMPER
(included). This jumper cable will connect the two outlets on the
back of the machine labeled UPS. Make sure to fully seat the
jumper cable.
4.3 Pressure Vessel Installation
The pressure vessel is shipped separately from the instrument; locate the pressure
vessel assembly and the motor/drive assembly. The pressure vessel is installed into
the heater block by carefully sliding it in from the top. A thermocouple is contained
inside the instrument.
The thermocouple must be inserted into the bottom sidewall of
the pressure vessel anytime the vessel is inserted into the
instrument.
Installation of the thermocouple may be easiest if the thermocouple is pulled
through the heater block and inserted before the pressure vessel is dropped into the
instrument.